Analysis of glycated hemoglobin A1c by capillary electrophoresis and capillary isoelectric focusing
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388963%3A_____%2F11%3A00359988" target="_blank" >RIV/61388963:_____/11:00359988 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.ab.2011.01.048" target="_blank" >http://dx.doi.org/10.1016/j.ab.2011.01.048</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.ab.2011.01.048" target="_blank" >10.1016/j.ab.2011.01.048</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Analysis of glycated hemoglobin A1c by capillary electrophoresis and capillary isoelectric focusing
Popis výsledku v původním jazyce
Two capillary electrophoretic methods were developed and evaluated for measurement of glycated hemoglobin HbA1c. Firstly, a capillary zone electrophoresis analysis is performed with a sodium tetraborate buffer of pH 9.3 as background electrolyte in a neutrally coated capillary. HbA1c is separated from HbA0 due to specific interactions of borate anions with the cis-diol pattern in the saccharide moiety of glycohemoglobin. Secondly, a capillary isoelectric focusing method, which exploits a difference in pI values of HbA0 and HbA1c, is performed with Servalyt pH 6-8 or alternatively with Biolyte pH 6-8 carrier ampholytes spiked with a narrow pH cut of pH 7.2 prepared by preparative fractionation of Servalyt pH 4-9 carrier ampholytes.
Název v anglickém jazyce
Analysis of glycated hemoglobin A1c by capillary electrophoresis and capillary isoelectric focusing
Popis výsledku anglicky
Two capillary electrophoretic methods were developed and evaluated for measurement of glycated hemoglobin HbA1c. Firstly, a capillary zone electrophoresis analysis is performed with a sodium tetraborate buffer of pH 9.3 as background electrolyte in a neutrally coated capillary. HbA1c is separated from HbA0 due to specific interactions of borate anions with the cis-diol pattern in the saccharide moiety of glycohemoglobin. Secondly, a capillary isoelectric focusing method, which exploits a difference in pI values of HbA0 and HbA1c, is performed with Servalyt pH 6-8 or alternatively with Biolyte pH 6-8 carrier ampholytes spiked with a narrow pH cut of pH 7.2 prepared by preparative fractionation of Servalyt pH 4-9 carrier ampholytes.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Biochemistry
ISSN
0003-2697
e-ISSN
—
Svazek periodika
413
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
8-15
Kód UT WoS článku
000289607900002