Pasivní spektrometrie přenosu lineární energie: vývoj a využití
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389005%3A_____%2F04%3A00100851" target="_blank" >RIV/61389005:_____/04:00100851 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Passive spectrometry of linear energy transfer: development and use
Popis výsledku v původním jazyce
A linear energy transfer (LET) spectrometer based on the evaluation of particle track parameters in a chemically etched polyallyldiglycolcarbonate (PADC) track detector has been developed at our laboratory. It permits us to determine LET spectra between10 and 700 keV ?mů1 in tissues. The LET spectra obtained permit us to calculate total dose and dose equivalent corresponding to particles with etchable tracks also. We have recently been able to verify the calibration curves used by using C, Mg, Ne, Si and Fe ion beams with different energies. The calibration curves obtained are presented and compared with those originally used, and a good correlation is found. The LET spectrometer with new calibration was used to analyse the radiation quality of the radiotherapy proton beam at the Joint Institute for Nuclear Research (JINR). The radiation quality was studied along the proton's range, particular attention being devoted to the region of the Bragg peak
Název v anglickém jazyce
Passive spectrometry of linear energy transfer: development and use
Popis výsledku anglicky
A linear energy transfer (LET) spectrometer based on the evaluation of particle track parameters in a chemically etched polyallyldiglycolcarbonate (PADC) track detector has been developed at our laboratory. It permits us to determine LET spectra between10 and 700 keV ?mů1 in tissues. The LET spectra obtained permit us to calculate total dose and dose equivalent corresponding to particles with etchable tracks also. We have recently been able to verify the calibration curves used by using C, Mg, Ne, Si and Fe ion beams with different energies. The calibration curves obtained are presented and compared with those originally used, and a good correlation is found. The LET spectrometer with new calibration was used to analyse the radiation quality of the radiotherapy proton beam at the Joint Institute for Nuclear Research (JINR). The radiation quality was studied along the proton's range, particular attention being devoted to the region of the Bragg peak
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Svazek periodika
110
Číslo periodika v rámci svazku
1-4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
675-679
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—