Proton beam dosimetry based on the graphene oxide reduction and Raman spectroscopy
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389005%3A_____%2F22%3A00561550" target="_blank" >RIV/61389005:_____/22:00561550 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1016/j.vacuum.2022.111113" target="_blank" >https://doi.org/10.1016/j.vacuum.2022.111113</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.vacuum.2022.111113" target="_blank" >10.1016/j.vacuum.2022.111113</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Proton beam dosimetry based on the graphene oxide reduction and Raman spectroscopy
Popis výsledku v původním jazyce
mu Raman spectra of graphene oxide (GO) and reduced graphene oxide (rGO) foils, subjected to 2 divided by 3 MeV proton beam irradiations in vacuum at low and high fluence, have been investigated. The spectra detailed analysis indicates that the characteristic D and G peaks yields ratio, the area of the D''peak, located between D and G, the G peak full width at half maximum, the C-H band area and full width at half maximum are functions of the absorbed dose. The D and the D'' peaks presence is indicative of the disorder generated in the target by the ion beam. All these observations suggest some useful indexes to read the proton absorbed dose by the irradiated rGO foils. Therefore, these GO based foils can be employed as dosimeters for proton beams in a large range of absorbed dose, from about 100 Gy up to about 114 MGy, within which Raman signals respond linearly.
Název v anglickém jazyce
Proton beam dosimetry based on the graphene oxide reduction and Raman spectroscopy
Popis výsledku anglicky
mu Raman spectra of graphene oxide (GO) and reduced graphene oxide (rGO) foils, subjected to 2 divided by 3 MeV proton beam irradiations in vacuum at low and high fluence, have been investigated. The spectra detailed analysis indicates that the characteristic D and G peaks yields ratio, the area of the D''peak, located between D and G, the G peak full width at half maximum, the C-H band area and full width at half maximum are functions of the absorbed dose. The D and the D'' peaks presence is indicative of the disorder generated in the target by the ion beam. All these observations suggest some useful indexes to read the proton absorbed dose by the irradiated rGO foils. Therefore, these GO based foils can be employed as dosimeters for proton beams in a large range of absorbed dose, from about 100 Gy up to about 114 MGy, within which Raman signals respond linearly.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_013%2F0001812" target="_blank" >EF16_013/0001812: Centrum urychlovačů a jaderných analytických metod - OP</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Vacuum
ISSN
0042-207X
e-ISSN
1879-2715
Svazek periodika
201
Číslo periodika v rámci svazku
JUL
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
111113
Kód UT WoS článku
000852716900001
EID výsledku v databázi Scopus
2-s2.0-85129852682