Quantification in MALDI-TOF mass spectrometry of modified polymers
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389013%3A_____%2F11%3A00359037" target="_blank" >RIV/61389013:_____/11:00359037 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.aca.2011.03.019" target="_blank" >http://dx.doi.org/10.1016/j.aca.2011.03.019</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2011.03.019" target="_blank" >10.1016/j.aca.2011.03.019</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Quantification in MALDI-TOF mass spectrometry of modified polymers
Popis výsledku v původním jazyce
MALDI-TOF MS quantification is hampered by the poor reproducibility and molecular-mass and compositional discrimination. The use of an internal standard increases reproducibility and allows a calibration curve to be constructed. With synthetic polymers,the method requires a polymer of the same molecular mass distribution and composition as the analyte. When the content of the precursor polymer in a modified polymer is sought, the pure precursor is usually available and the modified polymer can serve asan internal standard. The method has to be combined with that of standard addition and the precursor polymer added directly to the sample. The experiments with simulated modified polymers revealed the signal-intensity ratio not proportional to the concentrations ratio. Thus, the basic assumption of the standard-addition method is not met. Therefore, the multi-point combined internal-standard standard-addition method based on the power-type calibration curve was developed and verified.
Název v anglickém jazyce
Quantification in MALDI-TOF mass spectrometry of modified polymers
Popis výsledku anglicky
MALDI-TOF MS quantification is hampered by the poor reproducibility and molecular-mass and compositional discrimination. The use of an internal standard increases reproducibility and allows a calibration curve to be constructed. With synthetic polymers,the method requires a polymer of the same molecular mass distribution and composition as the analyte. When the content of the precursor polymer in a modified polymer is sought, the pure precursor is usually available and the modified polymer can serve asan internal standard. The method has to be combined with that of standard addition and the precursor polymer added directly to the sample. The experiments with simulated modified polymers revealed the signal-intensity ratio not proportional to the concentrations ratio. Thus, the basic assumption of the standard-addition method is not met. Therefore, the multi-point combined internal-standard standard-addition method based on the power-type calibration curve was developed and verified.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CC - Organická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA203%2F08%2F0543" target="_blank" >GA203/08/0543: Hybridní makromolekulární systémy na bázi hydrofilních polymerů a peptidů reagujících na vnější podněty</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Svazek periodika
693
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
7
Strana od-do
82-88
Kód UT WoS článku
000290886800010
EID výsledku v databázi Scopus
—