Comparison of digital holographic interferometry and constant temperature anemometry for measurement of temperature field in fluid
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389021%3A_____%2F15%3A00452981" target="_blank" >RIV/61389021:_____/15:00452981 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/46747885:24220/15:#0003365 RIV/46747885:24210/15:00002309
Výsledek na webu
<a href="http://dx.doi.org/10.1117/12.2185056" target="_blank" >http://dx.doi.org/10.1117/12.2185056</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1117/12.2185056" target="_blank" >10.1117/12.2185056</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of digital holographic interferometry and constant temperature anemometry for measurement of temperature field in fluid
Popis výsledku v původním jazyce
The presented paper shows possibility of using digital holographic interferometry (DHI) for temperature field measurement in moving fluids. This method uses a modified Twymann-Green setup having double sensitivity instead of commonly used Mach-Zehnder type of interferometer in order to obtain sufficient phases change of the field. On the other hand this setup is not light efficient as Mach-Zehnder interferometer. For measurement of the fast periodical phenomenon is not necessary to use always the high speed camera. One can consider this field to coherent phenomenon. With employing one digital camera synchronized to periodic field and external triggered one can capture whole period of the phenomenon. However the projections form one viewing direction ofasymmetrical temperature field maybe misguided. Hence for sufficient examination of the asymmetrical field one should capture a large number of the phenomenon?s projections from different viewing directions.
Název v anglickém jazyce
Comparison of digital holographic interferometry and constant temperature anemometry for measurement of temperature field in fluid
Popis výsledku anglicky
The presented paper shows possibility of using digital holographic interferometry (DHI) for temperature field measurement in moving fluids. This method uses a modified Twymann-Green setup having double sensitivity instead of commonly used Mach-Zehnder type of interferometer in order to obtain sufficient phases change of the field. On the other hand this setup is not light efficient as Mach-Zehnder interferometer. For measurement of the fast periodical phenomenon is not necessary to use always the high speed camera. One can consider this field to coherent phenomenon. With employing one digital camera synchronized to periodic field and external triggered one can capture whole period of the phenomenon. However the projections form one viewing direction ofasymmetrical temperature field maybe misguided. Hence for sufficient examination of the asymmetrical field one should capture a large number of the phenomenon?s projections from different viewing directions.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of SPIE Vol. 9508: Holography: Advances and Modern Trends IV
ISBN
978-1-62841-629-9
ISSN
0277-786X
e-ISSN
—
Počet stran výsledku
10
Strana od-do
—
Název nakladatele
SPIE-INT SOC OPTICAL ENGINEERING
Místo vydání
Bellingham
Místo konání akce
Prague
Datum konání akce
15. 4. 2015
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000357642900024