Evaluation of the WinROP system for identifying retinopathy of prematurity in czech preterm infants
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61988987%3A17110%2F17%3AA1801TXX" target="_blank" >RIV/61988987:17110/17:A1801TXX - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216224:14110/17:00096381 RIV/61989100:27740/17:10236935 RIV/00843989:_____/17:E0106118
Výsledek na webu
<a href="http://dx.doi.org/10.5507/bp.2016.061" target="_blank" >http://dx.doi.org/10.5507/bp.2016.061</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.5507/bp.2016.061" target="_blank" >10.5507/bp.2016.061</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Evaluation of the WinROP system for identifying retinopathy of prematurity in czech preterm infants
Popis výsledku v původním jazyce
Aims. Retinopathy of Prematurity (ROP) is a potentially serious condition that can afflict preterm infants. Timely and correct identification of individuals at risk of developing a serious form of ROP is therefore of paramount importance. WinROP is an online system for predicting ROP based on birth weight and weight increments. However, the results vary significantly for various populations. It has not been evaluated in the Czech population. This study evaluates the test characteristics (specificity, sensitivity, positive and negative predictive values) of the WinROP system in Czech preterm infants.Methods. Data on 445 prematurely born infants included in the ROP screening program at the University Hospital Ostrava, Czech Republic, were retrospectively entered into the WinROP system and the outcomes of the WinROP and regular screening were compared.Results. All 24 infants who developed high-risk (Type 1 or Type 2) ROP were correctly identified by the system. The sensitivity and negative predictive values for this group were 100%. However, the specificity and positive predictive values were substantially lower, resulting in a large number of false positives. Extending the analysis to low risk ROP, the system did not provide such reliable results.Conclusions. The system is a valuable tool for identifying infants who are not likely to develop high-risk ROP and this could help to substantially reduce the number of preterm infants in need of regular ROP screening. It is not suitable for predicting the development of less serious forms of ROP which is however in accordance with the declared aims of the WinROP system.
Název v anglickém jazyce
Evaluation of the WinROP system for identifying retinopathy of prematurity in czech preterm infants
Popis výsledku anglicky
Aims. Retinopathy of Prematurity (ROP) is a potentially serious condition that can afflict preterm infants. Timely and correct identification of individuals at risk of developing a serious form of ROP is therefore of paramount importance. WinROP is an online system for predicting ROP based on birth weight and weight increments. However, the results vary significantly for various populations. It has not been evaluated in the Czech population. This study evaluates the test characteristics (specificity, sensitivity, positive and negative predictive values) of the WinROP system in Czech preterm infants.Methods. Data on 445 prematurely born infants included in the ROP screening program at the University Hospital Ostrava, Czech Republic, were retrospectively entered into the WinROP system and the outcomes of the WinROP and regular screening were compared.Results. All 24 infants who developed high-risk (Type 1 or Type 2) ROP were correctly identified by the system. The sensitivity and negative predictive values for this group were 100%. However, the specificity and positive predictive values were substantially lower, resulting in a large number of false positives. Extending the analysis to low risk ROP, the system did not provide such reliable results.Conclusions. The system is a valuable tool for identifying infants who are not likely to develop high-risk ROP and this could help to substantially reduce the number of preterm infants in need of regular ROP screening. It is not suitable for predicting the development of less serious forms of ROP which is however in accordance with the declared aims of the WinROP system.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/LQ1602" target="_blank" >LQ1602: IT4Innovations excellence in science</a><br>
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
BIOMEDICAL PAPERS-OLOMOUC
ISSN
1213-8118
e-ISSN
1804-7521
Svazek periodika
1
Číslo periodika v rámci svazku
161
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
111-116
Kód UT WoS článku
000398093500017
EID výsledku v databázi Scopus
2-s2.0-85016285956