The effect of intake of water on the final values of body composition parameters in active athletes using two different bioimpedance analyzers
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61988987%3A17450%2F14%3AA1501C90" target="_blank" >RIV/61988987:17450/14:A1501C90 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The effect of intake of water on the final values of body composition parameters in active athletes using two different bioimpedance analyzers
Popis výsledku v původním jazyce
The objective of this study is to evaluate changes in the resulting values of body composition parameters as a consequence of water intake using two devices routinely applied in the field. The research was performed in a group of 33 soccer players (meanage 17.28?0.45 years). The measurements were taken using Tanita 418 - MA (frequency 50 kHz) and Nutriguard-M (frequency 100 kHz). The Tanita device showed statistically significant differences after the intake of 500 ml in parameters BH (+0.42 kg), BF (+0.39 kg, + 0.53%) and TBW (-0.38 %). As for the Nutriguard device, statistically significant differences were found in parameters TBW (+0.77 kg), ICW (+0.83 kg), FFM (+1.05 kg), BCM (+0.79 kg) and ECM/BCM (-0.01). Also, the water intake had no effect onthe accuracy of measurement of both devices. The measurements of the SE values before and after the water intake did not differ.
Název v anglickém jazyce
The effect of intake of water on the final values of body composition parameters in active athletes using two different bioimpedance analyzers
Popis výsledku anglicky
The objective of this study is to evaluate changes in the resulting values of body composition parameters as a consequence of water intake using two devices routinely applied in the field. The research was performed in a group of 33 soccer players (meanage 17.28?0.45 years). The measurements were taken using Tanita 418 - MA (frequency 50 kHz) and Nutriguard-M (frequency 100 kHz). The Tanita device showed statistically significant differences after the intake of 500 ml in parameters BH (+0.42 kg), BF (+0.39 kg, + 0.53%) and TBW (-0.38 %). As for the Nutriguard device, statistically significant differences were found in parameters TBW (+0.77 kg), ICW (+0.83 kg), FFM (+1.05 kg), BCM (+0.79 kg) and ECM/BCM (-0.01). Also, the water intake had no effect onthe accuracy of measurement of both devices. The measurements of the SE values before and after the water intake did not differ.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
AC - Archeologie, antropologie, etnologie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta gymnica
ISSN
2336-4912
e-ISSN
—
Svazek periodika
2
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
10
Strana od-do
107-116
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—