Biomechanics ? Problematic Loosening of Locking Screws from Plates.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27230%2F14%3A86090505" target="_blank" >RIV/61989100:27230/14:86090505 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Biomechanics ? Problematic Loosening of Locking Screws from Plates.
Popis výsledku v původním jazyce
Angularly stable plates have been used in medical and veterinary practice for the treatment of unstable fractures for several years. The principle is based on the fixation of a screw in the plate by means of the screw thread. The thread on the screw headmust have the same gradient as the thread on the shank; often the thread on the screw head is finer and has multiple starts. In the operating procedure, locking bone screws in the angularly stable plate are manually tightened using a torque clutch withtorque 1.5 Nm. Even if this procedure is complied with, when extracting the titanium alloy implant it is difficult (and sometimes impossible) to loosen some locking bone screws from the locking holes in the angularly stable plate. The results of the experiments led to the following recommendations: the torque should be reduced from 1.5 Nm to a lower value; conical (self-holding) screwdrivers should be used when inserting screws and cylindrical (non-self-holding) screwdrivers should be us
Název v anglickém jazyce
Biomechanics ? Problematic Loosening of Locking Screws from Plates.
Popis výsledku anglicky
Angularly stable plates have been used in medical and veterinary practice for the treatment of unstable fractures for several years. The principle is based on the fixation of a screw in the plate by means of the screw thread. The thread on the screw headmust have the same gradient as the thread on the shank; often the thread on the screw head is finer and has multiple starts. In the operating procedure, locking bone screws in the angularly stable plate are manually tightened using a torque clutch withtorque 1.5 Nm. Even if this procedure is complied with, when extracting the titanium alloy implant it is difficult (and sometimes impossible) to loosen some locking bone screws from the locking holes in the angularly stable plate. The results of the experiments led to the following recommendations: the torque should be reduced from 1.5 Nm to a lower value; conical (self-holding) screwdrivers should be used when inserting screws and cylindrical (non-self-holding) screwdrivers should be us
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JR - Ostatní strojírenství
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA03010804" target="_blank" >TA03010804: Osteosyntéza zlomenin nohy a ruky</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
New Methods of Damage and Failure Analysis of Structural Parts : book of abstracts : 6th international conference : September 8-11, 2014, Ostrava
ISBN
978-80-248-3488-7
ISSN
—
e-ISSN
—
Počet stran výsledku
2
Strana od-do
37-38
Název nakladatele
VŠB - Technická univerzita Ostrava
Místo vydání
Ostrava
Místo konání akce
Ostrava
Datum konání akce
8. 9. 2014
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—