Medial plate for treatment of distal tibia fractures
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27230%2F15%3A86094523" target="_blank" >RIV/61989100:27230/15:86094523 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61989100:27740/15:86094523 RIV/61988987:17110/15:A1701LOF RIV/00843989:_____/15:E0106128
Výsledek na webu
<a href="http://www.scopus.com/record/display.uri?view=basic&eid=2-s2.0-84945929979&origin=resultslist#references" target="_blank" >http://www.scopus.com/record/display.uri?view=basic&eid=2-s2.0-84945929979&origin=resultslist#references</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Medial plate for treatment of distal tibia fractures
Popis výsledku v původním jazyce
The objective of the present study was to carry out a strength and reliability assessment of a medial plate for the treatment of distal tibia fractures. This was performed via numerical modelling (FEM) and experimental methods (compression test and Electronic Speckle Pattern Interferometry - ESPI method). The plate is used for internal fixation in orthopaedics and traumatology. Analyses were performed for non-fused bone with comminuted fracture in the distal metaphysis (i.e. unsuccessful treatment, where all loads in the metaphysis are carried by the plate). An anatomical tibia model (based on CT images in Mimics software) was used for FE analysis. In the experiments, the bone was replaced with a shaped piece of spruce wood (based on Mimics software) and the plate was loaded until its failure (major plastic deformation). Numerical and experimental results were evaluated and compared.
Název v anglickém jazyce
Medial plate for treatment of distal tibia fractures
Popis výsledku anglicky
The objective of the present study was to carry out a strength and reliability assessment of a medial plate for the treatment of distal tibia fractures. This was performed via numerical modelling (FEM) and experimental methods (compression test and Electronic Speckle Pattern Interferometry - ESPI method). The plate is used for internal fixation in orthopaedics and traumatology. Analyses were performed for non-fused bone with comminuted fracture in the distal metaphysis (i.e. unsuccessful treatment, where all loads in the metaphysis are carried by the plate). An anatomical tibia model (based on CT images in Mimics software) was used for FE analysis. In the experiments, the bone was replaced with a shaped piece of spruce wood (based on Mimics software) and the plate was loaded until its failure (major plastic deformation). Numerical and experimental results were evaluated and compared.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JR - Ostatní strojírenství
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA03010804" target="_blank" >TA03010804: Osteosyntéza zlomenin nohy a ruky</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Experimental Stress Analysis 2015 : 53rd international conference : June 1-4, 2015, Český Krumlov, Czech Republic : book of extended abstracts
ISBN
978-80-01-05734-6
ISSN
—
e-ISSN
—
Počet stran výsledku
7
Strana od-do
431-437
Název nakladatele
Czech Technical University in Prague
Místo vydání
Praha
Místo konání akce
Český Krumlov
Datum konání akce
1. 6. 2015
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—