Invasive blood pressure curves simulation device
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F11%3A86085779" target="_blank" >RIV/61989100:27240/11:86085779 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/MECBME.2011.5752098" target="_blank" >http://dx.doi.org/10.1109/MECBME.2011.5752098</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/MECBME.2011.5752098" target="_blank" >10.1109/MECBME.2011.5752098</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Invasive blood pressure curves simulation device
Popis výsledku v původním jazyce
Patient monitor modules have various inputs for vital function measurement. We can practice many of these measurements with some students in the laboratory of biomedical engineering. However, invasive blood pressure (IBP) measurement demonstration is impossible. This paper is dealing with design and realization of a programmable invasive blood pressure simulator. This device is able to generate programmable behavior of voltage signal with correspond to blood pressure curve. The user communication interface of the device allows choosing the type of generated signal by LCD and 4 push buttons. The wide spectrums of generated signals, which correspond to physiological or pathological blood pressure curves, are stored in a user programmable memory. The output connectors of the device can be directly connected to a patient monitor IBP module input. Invasive blood pressure measurement simulation for educational purposes and calibrating the patients monitors are the general usages of the devel
Název v anglickém jazyce
Invasive blood pressure curves simulation device
Popis výsledku anglicky
Patient monitor modules have various inputs for vital function measurement. We can practice many of these measurements with some students in the laboratory of biomedical engineering. However, invasive blood pressure (IBP) measurement demonstration is impossible. This paper is dealing with design and realization of a programmable invasive blood pressure simulator. This device is able to generate programmable behavior of voltage signal with correspond to blood pressure curve. The user communication interface of the device allows choosing the type of generated signal by LCD and 4 push buttons. The wide spectrums of generated signals, which correspond to physiological or pathological blood pressure curves, are stored in a user programmable memory. The output connectors of the device can be directly connected to a patient monitor IBP module input. Invasive blood pressure measurement simulation for educational purposes and calibrating the patients monitors are the general usages of the devel
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
2011 1st Middle East Conference on Biomedical Engineering, MECBME 2011
ISBN
978-1-4244-7000-6
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
192-195
Název nakladatele
IEEE
Místo vydání
New York
Místo konání akce
Sharjah
Datum konání akce
21. 2. 2011
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—