Design and implementation of monitor tester for validation of ECG signals
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F19%3A10242735" target="_blank" >RIV/61989100:27240/19:10242735 - isvavai.cz</a>
Výsledek na webu
<a href="http://file:///C:/Users/kas73/Downloads/BIODEVICES_2019_2.pdf" target="_blank" >http://file:///C:/Users/kas73/Downloads/BIODEVICES_2019_2.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design and implementation of monitor tester for validation of ECG signals
Popis výsledku v původním jazyce
The paper deals with a design and consequent realization of a monitor tester for validation of the ECG devices, and patient's monitors for the ECG measurement. In the paper, we describe the electrical design of the plug-in module, which is intended for a transmission of the digital data from the computer. Data are sent by using the controlling SW connected to the parallel port in the plug-in module. An essential part of the proposed module is the three-channel 12 bit D/A converter AD 7398. This converter is able to simultaneously send data from all the three outputs. The converter output is unipolar in the volts units. For this reason, the transmitted signals are adjusted by using the differential amplifiers in order to convert them to the bipolar. The signal's amplitudes are decreased to the millivolts level by using the voltage dividers. On the base of this solution, the Einthoven ECG leads may be detected. (C) 2019 by SCITEPRESS - Science and Technology Publications, Lda.
Název v anglickém jazyce
Design and implementation of monitor tester for validation of ECG signals
Popis výsledku anglicky
The paper deals with a design and consequent realization of a monitor tester for validation of the ECG devices, and patient's monitors for the ECG measurement. In the paper, we describe the electrical design of the plug-in module, which is intended for a transmission of the digital data from the computer. Data are sent by using the controlling SW connected to the parallel port in the plug-in module. An essential part of the proposed module is the three-channel 12 bit D/A converter AD 7398. This converter is able to simultaneously send data from all the three outputs. The converter output is unipolar in the volts units. For this reason, the transmitted signals are adjusted by using the differential amplifiers in order to convert them to the bipolar. The signal's amplitudes are decreased to the millivolts level by using the voltage dividers. On the base of this solution, the Einthoven ECG leads may be detected. (C) 2019 by SCITEPRESS - Science and Technology Publications, Lda.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
30400 - Medical biotechnology
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
BIODEVICES 2019 - 12th International Conference on Biomedical Electronics and Devices, Proceedings; Part of 12th International Joint Conference on Biomedical Engineering Systems and Technologies, BIOSTEC 2019
ISBN
978-989-758-353-7
ISSN
—
e-ISSN
—
Počet stran výsledku
7
Strana od-do
127-133
Název nakladatele
SciTePress - Science and Technology Publications
Místo vydání
Setúbal
Místo konání akce
Praha
Datum konání akce
22. 2. 2019
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—