A Novel FBG-Based Triggering System for Cardiac MR Imaging at 3 Tesla: A Pilot Pre-Clinical Study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F20%3A10245658" target="_blank" >RIV/61989100:27240/20:10245658 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216224:14740/20:00118190
Výsledek na webu
<a href="https://ieeexplore.ieee.org/document/9210616" target="_blank" >https://ieeexplore.ieee.org/document/9210616</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/ACCESS.2020.3028224" target="_blank" >10.1109/ACCESS.2020.3028224</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A Novel FBG-Based Triggering System for Cardiac MR Imaging at 3 Tesla: A Pilot Pre-Clinical Study
Popis výsledku v původním jazyce
This first-ever study demonstrates the applicability of a fiber Bragg grating (FBG) system for MR cardiac triggering of cardiovascular magnetic resonance at 3 Tesla. The unique patented system senses body movements caused by cardiac activity using a non-invasive ballistocardiography (BCG) sensor. The pilot research compares a novel FBG-based system with clinically used triggering systems based on electrocardiography (ECG) and pulse oximetry (POX). The pilot pre-clinical study was conducted on 8 subjects at a Siemens Prisma 3T MR Scanner. The study compares images from two basic cardiac sequences, TRUE FISP (Free Induction Decay Steady-State Precession) and PSIR (Phase Sensitive Inversion Recovery), using objective methods and subjective evaluation by clinical experts. The study presents original results that confirm the applicability of optical sensors in the field of cardiac triggering having a number of advantages in comparison to conventional solutions, such as no eddy current interference, ease of placement of the sensor on the patient's body, and senor reusability. The proposed FBG-based system achieves comparable results with the most frequently used and most accurate ECG-based and POX-based clinical systems. In terms of subjective evaluation by experts, the FBG system outperformed the POX-based system used in clinical practice.
Název v anglickém jazyce
A Novel FBG-Based Triggering System for Cardiac MR Imaging at 3 Tesla: A Pilot Pre-Clinical Study
Popis výsledku anglicky
This first-ever study demonstrates the applicability of a fiber Bragg grating (FBG) system for MR cardiac triggering of cardiovascular magnetic resonance at 3 Tesla. The unique patented system senses body movements caused by cardiac activity using a non-invasive ballistocardiography (BCG) sensor. The pilot research compares a novel FBG-based system with clinically used triggering systems based on electrocardiography (ECG) and pulse oximetry (POX). The pilot pre-clinical study was conducted on 8 subjects at a Siemens Prisma 3T MR Scanner. The study compares images from two basic cardiac sequences, TRUE FISP (Free Induction Decay Steady-State Precession) and PSIR (Phase Sensitive Inversion Recovery), using objective methods and subjective evaluation by clinical experts. The study presents original results that confirm the applicability of optical sensors in the field of cardiac triggering having a number of advantages in comparison to conventional solutions, such as no eddy current interference, ease of placement of the sensor on the patient's body, and senor reusability. The proposed FBG-based system achieves comparable results with the most frequently used and most accurate ECG-based and POX-based clinical systems. In terms of subjective evaluation by experts, the FBG system outperformed the POX-based system used in clinical practice.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Access
ISSN
2169-3536
e-ISSN
—
Svazek periodika
8
Číslo periodika v rámci svazku
2020
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
19
Strana od-do
181205-181223
Kód UT WoS článku
000577884000001
EID výsledku v databázi Scopus
—