Potentiodynamic polarization and electrochemical impedance spectroscopy used for prediction of nitinol stent's lifetime
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27360%2F17%3A10238034" target="_blank" >RIV/61989100:27360/17:10238034 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.scientific.net/KEM.741.50" target="_blank" >https://www.scientific.net/KEM.741.50</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.4028/www.scientific.net/KEM.741.50" target="_blank" >10.4028/www.scientific.net/KEM.741.50</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Potentiodynamic polarization and electrochemical impedance spectroscopy used for prediction of nitinol stent's lifetime
Popis výsledku v původním jazyce
This paper is focused on prediction of nitinol stents lifetime. Stents have been tested before and after mechanical loading. Corrosion tests of mechanically loaded specimens indicated impairment of corrosion resistance and SEM observation confirmed presence of micro cracks in surface oxide layer created during heat treatment. Electrochemical impedance spectroscopy and potentiodynamic polarization measurements in physiological solution (0.9% NaCl in H2O) was used for these particular experiments. Corrosion rate and other parameters have been used for calculation of released nickel ions weight and prediction of application lifetime. © 2017 Trans Tech Publications, Switzerland.
Název v anglickém jazyce
Potentiodynamic polarization and electrochemical impedance spectroscopy used for prediction of nitinol stent's lifetime
Popis výsledku anglicky
This paper is focused on prediction of nitinol stents lifetime. Stents have been tested before and after mechanical loading. Corrosion tests of mechanically loaded specimens indicated impairment of corrosion resistance and SEM observation confirmed presence of micro cracks in surface oxide layer created during heat treatment. Electrochemical impedance spectroscopy and potentiodynamic polarization measurements in physiological solution (0.9% NaCl in H2O) was used for these particular experiments. Corrosion rate and other parameters have been used for calculation of released nickel ions weight and prediction of application lifetime. © 2017 Trans Tech Publications, Switzerland.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20501 - Materials engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Key Engineering Materials. Volume 741
ISBN
978-3-0357-1127-1
ISSN
1013-9826
e-ISSN
1662-9795
Počet stran výsledku
7
Strana od-do
50-56
Název nakladatele
Trans Tech Publications
Místo vydání
Uetikon-Zuerich
Místo konání akce
Yokohama
Datum konání akce
1. 11. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—