Stanovení fluoridů v moči iontově selektivní elektrodou: vnitrolaboratorní optimalizace
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15310%2F04%3A00002205" target="_blank" >RIV/61989592:15310/04:00002205 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61989592:15110/04:00000821 RIV/61989592:15110/04:00007231
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fluoride determination in urine with ion selective electrode: Within laboratory method and sample storage optimization
Popis výsledku v původním jazyce
This work presents method validation of fluoride determination in urine of children. The fluoride ion selective electrode and silver chloride reference electrode have been used for all measurements. We have found that described method is accurate (average recovery 98.4 %) and provided precision RSD = 2.8 %. The evaluated value of limit of detection using direct signal method has been 3.2.10-6 mol.l-1. The fluoride concentration in real urine samples is influenced during the storage in the refrigerator (+ 4 °C). The samples should be stored deeply frozen at - 20 °C before the analysis. The comparison of several buffers (commercially available TISAB, citrate buffer with different pH and citrate buffer with NaCl) leads to the conclusion that there is no significant difference among used buffer solutions.
Název v anglickém jazyce
Fluoride determination in urine with ion selective electrode: Within laboratory method and sample storage optimization
Popis výsledku anglicky
This work presents method validation of fluoride determination in urine of children. The fluoride ion selective electrode and silver chloride reference electrode have been used for all measurements. We have found that described method is accurate (average recovery 98.4 %) and provided precision RSD = 2.8 %. The evaluated value of limit of detection using direct signal method has been 3.2.10-6 mol.l-1. The fluoride concentration in real urine samples is influenced during the storage in the refrigerator (+ 4 °C). The samples should be stored deeply frozen at - 20 °C before the analysis. The comparison of several buffers (commercially available TISAB, citrate buffer with different pH and citrate buffer with NaCl) leads to the conclusion that there is no significant difference among used buffer solutions.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Universitatis Palackianae Olomucensis, Facultas Rerum Naturalium, Chemica
ISSN
0232-0061
e-ISSN
—
Svazek periodika
43
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
104-109
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—