Position Detection of an Analyte Phase in the Liquid Chromatography Column Using a Clamp-on Ultrasonic Monitor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62156489%3A43110%2F24%3A43926039" target="_blank" >RIV/62156489:43110/24:43926039 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.54937/2024.9788056111024.283-299" target="_blank" >https://doi.org/10.54937/2024.9788056111024.283-299</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.54937/2024.9788056111024.283-299" target="_blank" >10.54937/2024.9788056111024.283-299</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Position Detection of an Analyte Phase in the Liquid Chromatography Column Using a Clamp-on Ultrasonic Monitor
Popis výsledku v původním jazyce
In this study are results of an experimental analyte phase position detection in a liquid chromatograph column. To perform this experiment, a liquid chromatograph was used, with ethylacetate as the analyte and the ultrasonic portable flow monitor type Fluxus G608, primarily intended for measuring gas and liquid flows in industry. The experiment was based on measuring of different values of speed of sound during the movement of ethylacetate through the adsorbent column.The measurement results confirmed the possibility of using the given method to determine the position of the analyte in the chromatographic column.
Název v anglickém jazyce
Position Detection of an Analyte Phase in the Liquid Chromatography Column Using a Clamp-on Ultrasonic Monitor
Popis výsledku anglicky
In this study are results of an experimental analyte phase position detection in a liquid chromatograph column. To perform this experiment, a liquid chromatograph was used, with ethylacetate as the analyte and the ultrasonic portable flow monitor type Fluxus G608, primarily intended for measuring gas and liquid flows in industry. The experiment was based on measuring of different values of speed of sound during the movement of ethylacetate through the adsorbent column.The measurement results confirmed the possibility of using the given method to determine the position of the analyte in the chromatographic column.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
50204 - Business and management
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
DOKOR 2024: Recenzovaný zborník príspevkov z medzinárodnej interdisciplinárnej doktorandskej konferencie
ISBN
978-80-561-1102-4
ISSN
—
e-ISSN
—
Počet stran výsledku
17
Strana od-do
283-299
Název nakladatele
VERBUM - vydavateľstvo KU
Místo vydání
Ružomberok
Místo konání akce
Ružomberok
Datum konání akce
30. 4. 2024
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—