Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies
Popis výsledku
Identifikátory výsledku
Kód výsledku v IS VaVaI
Výsledek na webu
DOI - Digital Object Identifier
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies
Popis výsledku v původním jazyce
BackgroundThis study compares cardiac output (CO) measurements obtained by lithium dilution (LiDCO), pulse power analysis (PulseCO), and continuous thermodilution (CTD) with bolus thermodilution (BTD) in ponies.MethodsEight isoflurane-anaesthetized Shetland ponies received xylazine, ketamine, and midazolam infusions (0.3, 1.2, and 0.018 mg kg-1 h-1, respectively). CO was measured with BTD, CTD, LiDCO, and PulseCO. Lithium was injected into the jugular vein and blood was sampled from the facial artery for lithium detection and this artery was also used for PulseCO. Measurements were obtained during four stable haemodynamic conditions in the following order: isoflurane 1 (end-tidal concentration), isoflurane 2, isoflurane 1, and isoflurane 1dobutamine 5mg kg-1 min-1.ResultsThe bias (2 sd) was 2.5 (2.1) and 0.5 (2.9) litre min-1 for LiDCO-BTD and for CTD-BTD comparisons, respectively. The limits of agreement were wider than +-30; therefore, interchangeability was rejected for both compar
Název v anglickém jazyce
Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies
Popis výsledku anglicky
BackgroundThis study compares cardiac output (CO) measurements obtained by lithium dilution (LiDCO), pulse power analysis (PulseCO), and continuous thermodilution (CTD) with bolus thermodilution (BTD) in ponies.MethodsEight isoflurane-anaesthetized Shetland ponies received xylazine, ketamine, and midazolam infusions (0.3, 1.2, and 0.018 mg kg-1 h-1, respectively). CO was measured with BTD, CTD, LiDCO, and PulseCO. Lithium was injected into the jugular vein and blood was sampled from the facial artery for lithium detection and this artery was also used for PulseCO. Measurements were obtained during four stable haemodynamic conditions in the following order: isoflurane 1 (end-tidal concentration), isoflurane 2, isoflurane 1, and isoflurane 1dobutamine 5mg kg-1 min-1.ResultsThe bias (2 sd) was 2.5 (2.1) and 0.5 (2.9) litre min-1 for LiDCO-BTD and for CTD-BTD comparisons, respectively. The limits of agreement were wider than +-30; therefore, interchangeability was rejected for both compar
Klasifikace
Druh
Jx - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
GJ - Choroby a škůdci zvířat, veterinární medicina
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
British journal of anaesthesia
ISSN
0007-0912
e-ISSN
—
Svazek periodika
109
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
864-869
Kód UT WoS článku
000311304400005
EID výsledku v databázi Scopus
—
Základní informace
Druh výsledku
Jx - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP
GJ - Choroby a škůdci zvířat, veterinární medicina
Rok uplatnění
2012