Fatigue life of Drawn Filled NiTi-Ag Wires for Medical Applications
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62690094%3A18470%2F12%3A50000600" target="_blank" >RIV/62690094:18470/12:50000600 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68378297:_____/12:00381244
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fatigue life of Drawn Filled NiTi-Ag Wires for Medical Applications
Popis výsledku v původním jazyce
Nickel?titanium alloys (Nitinol) which possess the unique thermal shape memory and superelasticity effects are attracting more attention in medical devices. Currently it is mainly about the construction of stents or other types of implants. Researchers are looking for new ways to improve the mechanical and physical properties of materials for the manufacture of stents. New types of wires are tested. This new wires are prepared by drawn filled tubing technology. Drawn filled wires enable the unique ability to match dissimilar materials to provide a variety of properties in a single wire system. This technology can be utilized by the engineer to resolve technical issues. This article is devoted to fractographical analysis of fatigue fracture process in drawn filled wires made of nitinol with silver core and drawn filled wires made of silver with nitinol core.
Název v anglickém jazyce
Fatigue life of Drawn Filled NiTi-Ag Wires for Medical Applications
Popis výsledku anglicky
Nickel?titanium alloys (Nitinol) which possess the unique thermal shape memory and superelasticity effects are attracting more attention in medical devices. Currently it is mainly about the construction of stents or other types of implants. Researchers are looking for new ways to improve the mechanical and physical properties of materials for the manufacture of stents. New types of wires are tested. This new wires are prepared by drawn filled tubing technology. Drawn filled wires enable the unique ability to match dissimilar materials to provide a variety of properties in a single wire system. This technology can be utilized by the engineer to resolve technical issues. This article is devoted to fractographical analysis of fatigue fracture process in drawn filled wires made of nitinol with silver core and drawn filled wires made of silver with nitinol core.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JL - Únava materiálu a lomová mechanika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
DNCOCO 2012 : proceedings of the 11th WSEAS international conference on data networks, communications, computers
ISBN
978-1-61804-118-0
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
249-253
Název nakladatele
World scientific and engineering academy and society
Místo vydání
Athens
Místo konání akce
Sliema, Malta
Datum konání akce
7. 9. 2012
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—