Fully automated measurement of anisotropy of magnetic susceptibility using 3D rotator
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985831%3A_____%2F14%3A00432989" target="_blank" >RIV/67985831:_____/14:00432989 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.tecto.2014.02.015" target="_blank" >http://dx.doi.org/10.1016/j.tecto.2014.02.015</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.tecto.2014.02.015" target="_blank" >10.1016/j.tecto.2014.02.015</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fully automated measurement of anisotropy of magnetic susceptibility using 3D rotator
Popis výsledku v původním jazyce
A 3D rotator (for MFK1-FA or MFK1-A Kappabridges) was developed to increase the speed and comfort of anisotropy of magnetic susceptibility (AMS) measurements. The 3D rotator rotates the specimen simultaneously about two axes with different velocities. The two-axis rotation enables the determination of 320 directional susceptibilities during a single anisotropy measurement. These directions are very well distributed on a sphere which makes the measuring design almost rotatable. The actual measurement isfully automated in such a way that, once the specimen is mounted into the rotator, it requires no additional manipulation to measure the full AMS tensor. The approximate duration of one anisotropy measurement is 1.5 min. Fundamental ideas of data acquisition and processing, AMS tensor fitting and respective error analysis are described and discussed. Calculation of the anisotropy tensor, respective error analysis and immediate data visualization is provided by Safyr5 software. Examples c
Název v anglickém jazyce
Fully automated measurement of anisotropy of magnetic susceptibility using 3D rotator
Popis výsledku anglicky
A 3D rotator (for MFK1-FA or MFK1-A Kappabridges) was developed to increase the speed and comfort of anisotropy of magnetic susceptibility (AMS) measurements. The 3D rotator rotates the specimen simultaneously about two axes with different velocities. The two-axis rotation enables the determination of 320 directional susceptibilities during a single anisotropy measurement. These directions are very well distributed on a sphere which makes the measuring design almost rotatable. The actual measurement isfully automated in such a way that, once the specimen is mounted into the rotator, it requires no additional manipulation to measure the full AMS tensor. The approximate duration of one anisotropy measurement is 1.5 min. Fundamental ideas of data acquisition and processing, AMS tensor fitting and respective error analysis are described and discussed. Calculation of the anisotropy tensor, respective error analysis and immediate data visualization is provided by Safyr5 software. Examples c
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DE - Zemský magnetismus, geodesie, geografie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Tectonophysics
ISSN
0040-1951
e-ISSN
—
Svazek periodika
629
Číslo periodika v rámci svazku
26 August
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
8
Strana od-do
6-13
Kód UT WoS článku
000341474600002
EID výsledku v databázi Scopus
—