Comprehensive Characterization of Original 10-stage and 7-stage Modified Berner Type Impactors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985858%3A_____%2F11%3A00353090" target="_blank" >RIV/67985858:_____/11:00353090 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comprehensive Characterization of Original 10-stage and 7-stage Modified Berner Type Impactors
Popis výsledku v původním jazyce
Two cascade impactors, Berner Low Pressure Impactor and modified BLPI (with lower pressure drop through) were characterized experimentally. First, the critical dimensions affecting the collection characteristics were measured for each impactor stage. Second, the stage absolute pressures were determined. Finally, the collection efficiency curves were obtained for all stages covering the size range of 0.026 ? 13.7 ?m. The collection efficiency of modified impactor was measured using two condensation particle counters (CPC). In addition, both impactors were experimentally tested for their particle losses using the ammonium sulfate particles at dry (15% RH) and wet (>80% RH) conditions. The losses were determined by rinsing sampling foil and correspondingimpactor stage (jet plate, walls and spacer ring) by deionized water. Subsequently the ammonium sulphate content in a sample together with rinse water was analyzed using ion chromatography.
Název v anglickém jazyce
Comprehensive Characterization of Original 10-stage and 7-stage Modified Berner Type Impactors
Popis výsledku anglicky
Two cascade impactors, Berner Low Pressure Impactor and modified BLPI (with lower pressure drop through) were characterized experimentally. First, the critical dimensions affecting the collection characteristics were measured for each impactor stage. Second, the stage absolute pressures were determined. Finally, the collection efficiency curves were obtained for all stages covering the size range of 0.026 ? 13.7 ?m. The collection efficiency of modified impactor was measured using two condensation particle counters (CPC). In addition, both impactors were experimentally tested for their particle losses using the ammonium sulfate particles at dry (15% RH) and wet (>80% RH) conditions. The losses were determined by rinsing sampling foil and correspondingimpactor stage (jet plate, walls and spacer ring) by deionized water. Subsequently the ammonium sulphate content in a sample together with rinse water was analyzed using ion chromatography.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Aerosol Science and Technology
ISSN
0278-6826
e-ISSN
—
Svazek periodika
45
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
13
Strana od-do
—
Kód UT WoS článku
000285542900009
EID výsledku v databázi Scopus
—