Fiber-optic pH sensing system with microscopic spatial resolution
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985882%3A_____%2F15%3A00457542" target="_blank" >RIV/67985882:_____/15:00457542 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1117/12.2179471" target="_blank" >http://dx.doi.org/10.1117/12.2179471</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1117/12.2179471" target="_blank" >10.1117/12.2179471</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fiber-optic pH sensing system with microscopic spatial resolution
Popis výsledku v původním jazyce
Acidity of an environment expressed as pH is one of the key biological parameters. An optical sensing scheme together with the use of optical fiber probes can bring advantages to the pH monitoring such as fine spatial resolution, less interference with biological material or immunity to the electromagnetic field. An optical pH-sensor with fine spatial resolution suitable for measurement of pH in near-neutral range is presented and the influence of sensing probe preparation process onto the sensor performance is discussed in the paper. The sensor is based on an ion pair of 8- hydroxypyrrene-1,3,6-pyrene trisulfonic acid trisodium salt (HPTS) and hexadecyltrimethylammonium bromide (CTAB) immobilized by a sol-gel method on the tip of tapered-fiber probes.It was found that the sensor performance is affected significantly by the shape of the sensing layer
Název v anglickém jazyce
Fiber-optic pH sensing system with microscopic spatial resolution
Popis výsledku anglicky
Acidity of an environment expressed as pH is one of the key biological parameters. An optical sensing scheme together with the use of optical fiber probes can bring advantages to the pH monitoring such as fine spatial resolution, less interference with biological material or immunity to the electromagnetic field. An optical pH-sensor with fine spatial resolution suitable for measurement of pH in near-neutral range is presented and the influence of sensing probe preparation process onto the sensor performance is discussed in the paper. The sensor is based on an ion pair of 8- hydroxypyrrene-1,3,6-pyrene trisulfonic acid trisodium salt (HPTS) and hexadecyltrimethylammonium bromide (CTAB) immobilized by a sol-gel method on the tip of tapered-fiber probes.It was found that the sensor performance is affected significantly by the shape of the sensing layer
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA04011400" target="_blank" >TA04011400: Vláknově optický pH metr pro mikroskopické biologické vzorky</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Optical Sensors 2015 (Proceedings of SPIE 9506)
ISBN
9781628416275
ISSN
0277-786X
e-ISSN
—
Počet stran výsledku
6
Strana od-do
9506121-9506126
Název nakladatele
SPIE
Místo vydání
Bellingham
Místo konání akce
Prague
Datum konání akce
13. 4. 2015
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000356702200025