Perchlorate Solid-Contact Ion-Selective Electrode Based on Dodecabenzylbambus[6]uril
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081715%3A_____%2F22%3A00556651" target="_blank" >RIV/68081715:_____/22:00556651 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216224:14310/22:00125586
Výsledek na webu
<a href="https://www.mdpi.com/2227-9040/10/3/115" target="_blank" >https://www.mdpi.com/2227-9040/10/3/115</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/chemosensors10030115" target="_blank" >10.3390/chemosensors10030115</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Perchlorate Solid-Contact Ion-Selective Electrode Based on Dodecabenzylbambus[6]uril
Popis výsledku v původním jazyce
Dodecabenzylbambus[6]uril (Bn12BU[6]) is an anion receptor that binds the perchlorate ion the most tightly (stability constant similar to 1010 M-1) of all anions due to the excellent match between the ion size in relation to the receptor cavity. This new bambusuril compound was used as an ionophore in the ion-selective membrane (ISM) to develop ion selective electrodes (ISEs) for determination of perchlorate concentration utilizing the poly(3,4-ethylenedioxythiophene) (PEDOT) polymer film as a solid-contact material. Variation of the content of Bn12BU[6] and tridodecylmethylammonium chloride (TDMAC1) in the plasticized poly(vinyl chloride)-based ISM was also tested. All the prepared solid-contact ISEs and their analytical performance were characterized by potentiometry, cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and chronopotentiometry. The ISEs showed rapid response and a sub-Nernstian slope (similar to 57 mV/decade) during potentiometric measurements in perchlorate solutions in the concentration range from 10-1 to 10-6 M simultaneously with their high stability and sufficient selectivity to other common inorganic anions like bromide, chloride, nitrate and sulphate. The function of the ISE was further verified by analysis of real water samples (lake, sea, and mineral water), which gave accurate and precise results.
Název v anglickém jazyce
Perchlorate Solid-Contact Ion-Selective Electrode Based on Dodecabenzylbambus[6]uril
Popis výsledku anglicky
Dodecabenzylbambus[6]uril (Bn12BU[6]) is an anion receptor that binds the perchlorate ion the most tightly (stability constant similar to 1010 M-1) of all anions due to the excellent match between the ion size in relation to the receptor cavity. This new bambusuril compound was used as an ionophore in the ion-selective membrane (ISM) to develop ion selective electrodes (ISEs) for determination of perchlorate concentration utilizing the poly(3,4-ethylenedioxythiophene) (PEDOT) polymer film as a solid-contact material. Variation of the content of Bn12BU[6] and tridodecylmethylammonium chloride (TDMAC1) in the plasticized poly(vinyl chloride)-based ISM was also tested. All the prepared solid-contact ISEs and their analytical performance were characterized by potentiometry, cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and chronopotentiometry. The ISEs showed rapid response and a sub-Nernstian slope (similar to 57 mV/decade) during potentiometric measurements in perchlorate solutions in the concentration range from 10-1 to 10-6 M simultaneously with their high stability and sufficient selectivity to other common inorganic anions like bromide, chloride, nitrate and sulphate. The function of the ISE was further verified by analysis of real water samples (lake, sea, and mineral water), which gave accurate and precise results.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/LTC20044" target="_blank" >LTC20044: Thermodynamic and kinetic investigations of metal complexes of polydentate ligands for biomedical, analytical and industrial applications</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemosensors
ISSN
2227-9040
e-ISSN
2227-9040
Svazek periodika
10
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
16
Strana od-do
115
Kód UT WoS článku
000776842000001
EID výsledku v databázi Scopus
2-s2.0-85127634761