Zařízení pro měření tepelné emisivity při nízkých teplotách
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F04%3A00100016" target="_blank" >RIV/68081731:_____/04:00100016 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Device for measurement of thermal emissivity at cryogenics temperatures
Popis výsledku v původním jazyce
In the described device, the thermal emissivity or absorptivity of the sample is measured by substitution of the radiative heat flow between two parallel surfaces by thermal output of a heater. Fast measurements of the mutual emissivity for the range ofthe temperature of the radiating surface 25 K - 150 K are possible. The absorbing surface has a temperature between 5 K and 10 K when LHe is used as cryoliquid. The desired measurement sensitivity is 1 mK for temperature and 0.1 mikroW for heat power, respectively. The diameter of the whole device is 50 mm and so it is possible to use a commercial Dewar can for the cooling. The form of the sample is a circular plate 40 mm in diameter and 1 mm in thickness with one tested side. The emissivity and its temperature dependence for various surface treatments can be checked immediately before application in a cryogenic system
Název v anglickém jazyce
Device for measurement of thermal emissivity at cryogenics temperatures
Popis výsledku anglicky
In the described device, the thermal emissivity or absorptivity of the sample is measured by substitution of the radiative heat flow between two parallel surfaces by thermal output of a heater. Fast measurements of the mutual emissivity for the range ofthe temperature of the radiating surface 25 K - 150 K are possible. The absorbing surface has a temperature between 5 K and 10 K when LHe is used as cryoliquid. The desired measurement sensitivity is 1 mK for temperature and 0.1 mikroW for heat power, respectively. The diameter of the whole device is 50 mm and so it is possible to use a commercial Dewar can for the cooling. The form of the sample is a circular plate 40 mm in diameter and 1 mm in thickness with one tested side. The emissivity and its temperature dependence for various surface treatments can be checked immediately before application in a cryogenic system
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BJ - Termodynamika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/IBS2065109" target="_blank" >IBS2065109: Materiály pro kryotechniku a měření jejich emisivit</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 8th Cryogenics 2004 IIR international conference
ISBN
2-913149-33-2
ISSN
—
e-ISSN
—
Počet stran výsledku
7
Strana od-do
23-29
Název nakladatele
ICARIS
Místo vydání
Praha
Místo konání akce
Praha
Datum konání akce
27. 4. 2004
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—