Elimination of the Respiratory Effect on the Thoracic Impedance Signal with Whole-body Impedance Cardiography
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F10%3A00441035" target="_blank" >RIV/68081731:_____/10:00441035 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Elimination of the Respiratory Effect on the Thoracic Impedance Signal with Whole-body Impedance Cardiography
Popis výsledku v původním jazyce
The study presents a method for quantification of the respiratory effect on the thoracic impedance signal. We used measurements with 18 simultaneously scanning impedance signal locations. Elimination of breathing is based on the assumption that the neckand limb impedance signal measures only the variability in systemic blood pressure and is not affected by respiration. The results show approximately double the respiration variability in thorax impedance parameters in comparison with legs and neck. Conclusion: Whole-body impedance signal measurements and processing reflect changes in the hemodynamic system not affected by respiration. By using retrospective reconstruction we can eliminate the influence of respiration on the impedance signal from the chest.
Název v anglickém jazyce
Elimination of the Respiratory Effect on the Thoracic Impedance Signal with Whole-body Impedance Cardiography
Popis výsledku anglicky
The study presents a method for quantification of the respiratory effect on the thoracic impedance signal. We used measurements with 18 simultaneously scanning impedance signal locations. Elimination of breathing is based on the assumption that the neckand limb impedance signal measures only the variability in systemic blood pressure and is not affected by respiration. The results show approximately double the respiration variability in thorax impedance parameters in comparison with legs and neck. Conclusion: Whole-body impedance signal measurements and processing reflect changes in the hemodynamic system not affected by respiration. By using retrospective reconstruction we can eliminate the influence of respiration on the impedance signal from the chest.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/IAA200650801" target="_blank" >IAA200650801: Selektivní impedanční kardiografie pro přesné stanovení hemodynamických parametrů krevní cirkulace</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Computing in Cardiology 2010
ISBN
978-1-4244-7318-2
ISSN
0276-6574
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1051-1054
Název nakladatele
IEEE
Místo vydání
New York
Místo konání akce
Belfast
Datum konání akce
26. 9. 2010
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000328956000264