Modeling and Measuring of the Changes of the Magnetic Field Next to Non-Ferromagnetic Substances for Magnetic Susceptibility Calculation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F11%3A00368262" target="_blank" >RIV/68081731:_____/11:00368262 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216305:26220/11:PU92761
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Modeling and Measuring of the Changes of the Magnetic Field Next to Non-Ferromagnetic Substances for Magnetic Susceptibility Calculation
Popis výsledku v původním jazyce
This article is focused on the principles of the magnetic susceptibility measurement of the non-ferromagnetic substances using NMR tomography. Magnetic susceptibility is calculated from changes of the magnetic field close to the cylinder shaped specimen.COMSOL program was used for modeling experiment results. Consequently, model results were compared to MR tomograph measured data. Changes of the magnetic field in space of xy cut of the specimen were integrated in both cases. Before integrating of the magnetic field changes was necessary to filter, unwrap and detect accurate position of the specimen. Magnetic susceptibility calculated from measured data is in comparison to theoretical value, as well as model value, slightly different. Difference of those values is lower than 5%. Due to the used model and measurement were extended principles of the NMR method for magnetic susceptibility measurement.
Název v anglickém jazyce
Modeling and Measuring of the Changes of the Magnetic Field Next to Non-Ferromagnetic Substances for Magnetic Susceptibility Calculation
Popis výsledku anglicky
This article is focused on the principles of the magnetic susceptibility measurement of the non-ferromagnetic substances using NMR tomography. Magnetic susceptibility is calculated from changes of the magnetic field close to the cylinder shaped specimen.COMSOL program was used for modeling experiment results. Consequently, model results were compared to MR tomograph measured data. Changes of the magnetic field in space of xy cut of the specimen were integrated in both cases. Before integrating of the magnetic field changes was necessary to filter, unwrap and detect accurate position of the specimen. Magnetic susceptibility calculated from measured data is in comparison to theoretical value, as well as model value, slightly different. Difference of those values is lower than 5%. Due to the used model and measurement were extended principles of the NMR method for magnetic susceptibility measurement.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Measurement 2011
ISBN
978-80-969672-4-7
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
213-216
Název nakladatele
Slovak Academy of Sciences
Místo vydání
Bratislava
Místo konání akce
Smolenice
Datum konání akce
27. 4. 2011
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—