Self-referenced interferometer for form measurement of hollow cylinders
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F16%3A00468514" target="_blank" >RIV/68081731:_____/16:00468514 - isvavai.cz</a>
Výsledek na webu
<a href="http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2595406&resultClick=1" target="_blank" >http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2595406&resultClick=1</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1117/12.2262989" target="_blank" >10.1117/12.2262989</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Self-referenced interferometer for form measurement of hollow cylinders
Popis výsledku v původním jazyce
We present a new self-referenced interferometer for the form measurement of hollow cylindrical tubes that provides the accuracy in the micrometer range. It is based on a simple and robust setup where the reference and object waves are represented by the central and peripheral parts, respectively, of the conical wave generated by a single axicon. The inteferogram is characterized by a closed-fringe pattern with a circular carrier which allows for the interference phase demodulation using spatial synchronous detection. The misalignment aberrations are corrected by an iterative computational procedure. The measurement capabilities of the interferometer are experimentally tested for the glass and stainless steel tubes with lengths from 220 mm to 600 mm.
Název v anglickém jazyce
Self-referenced interferometer for form measurement of hollow cylinders
Popis výsledku anglicky
We present a new self-referenced interferometer for the form measurement of hollow cylindrical tubes that provides the accuracy in the micrometer range. It is based on a simple and robust setup where the reference and object waves are represented by the central and peripheral parts, respectively, of the conical wave generated by a single axicon. The inteferogram is characterized by a closed-fringe pattern with a circular carrier which allows for the interference phase demodulation using spatial synchronous detection. The misalignment aberrations are corrected by an iterative computational procedure. The measurement capabilities of the interferometer are experimentally tested for the glass and stainless steel tubes with lengths from 220 mm to 600 mm.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
20th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics (Proceedings of SPIE 10142)
ISBN
—
ISSN
0277-786X
e-ISSN
—
Počet stran výsledku
6
Strana od-do
1-5
Název nakladatele
SPIE
Místo vydání
Bellingham
Místo konání akce
Jasná
Datum konání akce
5. 9. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000393152700069