Automatic system for gauge block calibration optimized to meet legal length metrology requirements
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F16%3A00468526" target="_blank" >RIV/68081731:_____/16:00468526 - isvavai.cz</a>
Výsledek na webu
<a href="http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2595360" target="_blank" >http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2595360</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1117/12.2263367" target="_blank" >10.1117/12.2263367</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Automatic system for gauge block calibration optimized to meet legal length metrology requirements
Popis výsledku v původním jazyce
This paper presents a contactless method for gauge blocks calibration combining laser interferometry and low-coherence interferometry. In the presented system, the contactless measurement of the absolute gauge block length is done as a single-step operation without any change in optical setup during measurement, giving complete information about the gauge block length. The paper also presents a set of optimization steps which have been done in order to transform the original experimental setup into the automatic system which meets legal length metrology requirements. To prove the measurement traceability, we conducted a set of gauge block length measurement comparing data from the optimized system and the established reference systems TESA NPL A.G.I. 300 and TESA–UPC operated in Czech Metrology Institute laboratory.
Název v anglickém jazyce
Automatic system for gauge block calibration optimized to meet legal length metrology requirements
Popis výsledku anglicky
This paper presents a contactless method for gauge blocks calibration combining laser interferometry and low-coherence interferometry. In the presented system, the contactless measurement of the absolute gauge block length is done as a single-step operation without any change in optical setup during measurement, giving complete information about the gauge block length. The paper also presents a set of optimization steps which have been done in order to transform the original experimental setup into the automatic system which meets legal length metrology requirements. To prove the measurement traceability, we conducted a set of gauge block length measurement comparing data from the optimized system and the established reference systems TESA NPL A.G.I. 300 and TESA–UPC operated in Czech Metrology Institute laboratory.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
20th Slovak-Czech-Polish Optical Conference on Wave and Quantum Aspects of Contemporary Optics (Proceedings of SPIE 10142)
ISBN
—
ISSN
0277-786X
e-ISSN
—
Počet stran výsledku
10
Strana od-do
1-10
Název nakladatele
SPIE
Místo vydání
Bellingham
Místo konání akce
Jasná
Datum konání akce
5. 9. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000393152700023