The VED meter - A new tool to measure the ventricular conduction abnormalities in heart failure patients
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F17%3A00487040" target="_blank" >RIV/68081731:_____/17:00487040 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.cinc.org/archives/2017/" target="_blank" >http://www.cinc.org/archives/2017/</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.22489/CinC.2017.377-059" target="_blank" >10.22489/CinC.2017.377-059</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The VED meter - A new tool to measure the ventricular conduction abnormalities in heart failure patients
Popis výsledku v původním jazyce
Heart failure patients with a left ventricular ejection fraction below or equal to 35%, QRS duration above 150 ms, and left bundle branch block (LBBB) are usually recommended for treatment with a cardiac resynchronization therapy (CRT) device. Yet, about one-third of these patients do not respond to CRT revealing the limit of the current patient management guidelines. For this reason, new methods for assessing the level of electrical ventricular delay (VED) are being explored. In this work, we present the VED Meter, autonomous software measuring the level of electrical activation delay from the standard body-surface electrocardiogram (ECG). The software delivers the VED parameter expressed in ms that reflects the electrical dyssynchrony between the right ventricle, septum and left ventricle lateral wall. We applied this software to the data from the MADIT-CRT trial to evaluate the VED values in three types of ventricular conduction disturbances: LBBB, right bundle branch block (RBBB), and intra-ventricular conduction disturbances (IVCD). We report that VED parameter has negative values for RBBB (delayed activation of the septum and the right ventricle) and positive for LBBB (delayed activation of the left ventricle lateral wall) which were significantly different at p<0.001.
Název v anglickém jazyce
The VED meter - A new tool to measure the ventricular conduction abnormalities in heart failure patients
Popis výsledku anglicky
Heart failure patients with a left ventricular ejection fraction below or equal to 35%, QRS duration above 150 ms, and left bundle branch block (LBBB) are usually recommended for treatment with a cardiac resynchronization therapy (CRT) device. Yet, about one-third of these patients do not respond to CRT revealing the limit of the current patient management guidelines. For this reason, new methods for assessing the level of electrical ventricular delay (VED) are being explored. In this work, we present the VED Meter, autonomous software measuring the level of electrical activation delay from the standard body-surface electrocardiogram (ECG). The software delivers the VED parameter expressed in ms that reflects the electrical dyssynchrony between the right ventricle, septum and left ventricle lateral wall. We applied this software to the data from the MADIT-CRT trial to evaluate the VED values in three types of ventricular conduction disturbances: LBBB, right bundle branch block (RBBB), and intra-ventricular conduction disturbances (IVCD). We report that VED parameter has negative values for RBBB (delayed activation of the septum and the right ventricle) and positive for LBBB (delayed activation of the left ventricle lateral wall) which were significantly different at p<0.001.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20602 - Medical laboratory technology (including laboratory samples analysis; diagnostic technologies) (Biomaterials to be 2.9 [physical characteristics of living material as related to medical implants, devices, sensors])
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Computing in Cardiology 2017
ISBN
—
ISSN
2325-887X
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1-4
Název nakladatele
Computing in Cardiology
Místo vydání
Rennes
Místo konání akce
Rennes
Datum konání akce
24. 9. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000450651100328