Automatic detection of P wave in ECG during ventricular extrasystoles
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F19%3A00494537" target="_blank" >RIV/68081731:_____/19:00494537 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1007/978-981-10-9038-7_72" target="_blank" >http://dx.doi.org/10.1007/978-981-10-9038-7_72</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-981-10-9038-7_72" target="_blank" >10.1007/978-981-10-9038-7_72</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Automatic detection of P wave in ECG during ventricular extrasystoles
Popis výsledku v původním jazyce
This work introduces a new method for P wave detection in ECG signals during ventricular extrasystoles. The authors of previous works which deal with detection of P waves tested their algorithms mainly on physiological records (sinus rhythm) and they reached good results for these records. Testing of P wave detection algorithms using pathological records is usually not provided and if it is, the results are notably worse than in the case of physiological records. The automatic and reliable detection of atrial activity in pathological situations is still an unsolved problem. In this work, phasor transform in combination with classification algorithm is used for P wave detection. Phasor transform converts each ECG sample into a phasor which enhances changes in the ECG signal. The classification is based on extraction of morphological features which are derived from each QRS complex. The results of classification are used for demarcation of areas in which P waves are searched using phasor transform. The proposed algorithm was tested on signals no. 106, 119, 214 and 223 from MIT-BIH arrhythmia database, in which the ventricular extrasystoles are present. For validation whether the algorithm is functional also for signals with physiological rhythm, it was tested on the signals no. 100, 101, 103, 117, and 122. The accuracy of the P wave detection in signals with ventricular extrasystoles is Se = 98.94% and PP = 98.30% and in signals without pathology is Se = 98.47% and PP = 99.99%.
Název v anglickém jazyce
Automatic detection of P wave in ECG during ventricular extrasystoles
Popis výsledku anglicky
This work introduces a new method for P wave detection in ECG signals during ventricular extrasystoles. The authors of previous works which deal with detection of P waves tested their algorithms mainly on physiological records (sinus rhythm) and they reached good results for these records. Testing of P wave detection algorithms using pathological records is usually not provided and if it is, the results are notably worse than in the case of physiological records. The automatic and reliable detection of atrial activity in pathological situations is still an unsolved problem. In this work, phasor transform in combination with classification algorithm is used for P wave detection. Phasor transform converts each ECG sample into a phasor which enhances changes in the ECG signal. The classification is based on extraction of morphological features which are derived from each QRS complex. The results of classification are used for demarcation of areas in which P waves are searched using phasor transform. The proposed algorithm was tested on signals no. 106, 119, 214 and 223 from MIT-BIH arrhythmia database, in which the ventricular extrasystoles are present. For validation whether the algorithm is functional also for signals with physiological rhythm, it was tested on the signals no. 100, 101, 103, 117, and 122. The accuracy of the P wave detection in signals with ventricular extrasystoles is Se = 98.94% and PP = 98.30% and in signals without pathology is Se = 98.47% and PP = 99.99%.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
IFMBE Proceedings
ISBN
978-981-10-9037-0
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
5
Strana od-do
381-385
Název nakladatele
Springer
Místo vydání
Singapore
Místo konání akce
Prague
Datum konání akce
3. 6. 2018
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000449742700072