SUPER-RESOLUTION LIGHT MICROSCOPY: STED microscope fine laser co-alignment influences the final resolution and SNR
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378050%3A_____%2F19%3A00523243" target="_blank" >RIV/68378050:_____/19:00523243 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
SUPER-RESOLUTION LIGHT MICROSCOPY: STED microscope fine laser co-alignment influences the final resolution and SNR
Popis výsledku v původním jazyce
The super-resolution (SR) methods are rapidly developing approaches in light microscopy field. The Stimulated emission depletion (STED) microscopy is SR method based on confocal scanning microscope. During the acquisition scanning, two lasers – excitation and depletion – are co-aligned into one single spot in the sample. For the maximal efficiency of the SR method, the proper co-alignment of used laser beams is crucial. The article explains the STED microscope operator recommended routine for the alignment. Simple experiments demonstrate significantly decreased number of incoming photons in the misaligned system. It naturally leads to the lower resolution during the acquisition and these important aspects are discussed within the article.
Název v anglickém jazyce
SUPER-RESOLUTION LIGHT MICROSCOPY: STED microscope fine laser co-alignment influences the final resolution and SNR
Popis výsledku anglicky
The super-resolution (SR) methods are rapidly developing approaches in light microscopy field. The Stimulated emission depletion (STED) microscopy is SR method based on confocal scanning microscope. During the acquisition scanning, two lasers – excitation and depletion – are co-aligned into one single spot in the sample. For the maximal efficiency of the SR method, the proper co-alignment of used laser beams is crucial. The article explains the STED microscope operator recommended routine for the alignment. Simple experiments demonstrate significantly decreased number of incoming photons in the misaligned system. It naturally leads to the lower resolution during the acquisition and these important aspects are discussed within the article.
Klasifikace
Druh
C - Kapitola v odborné knize
CEP obor
—
OECD FORD obor
10601 - Cell biology
Návaznosti výsledku
Projekt
<a href="/cs/project/LO1419" target="_blank" >LO1419: Biomodely pro zdraví - Centrum modelových organismů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název knihy nebo sborníku
Advances in Disease Models
ISBN
978-80-88011-06-4
Počet stran výsledku
7
Strana od-do
234-240
Počet stran knihy
310
Název nakladatele
OPTIO CZ
Místo vydání
Praha
Kód UT WoS kapitoly
—