Corrosion and mechanical properties of atomic layer deposited TiO2 Coatings on NiTi implants
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F16%3A00486501" target="_blank" >RIV/68378271:_____/16:00486501 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/ICIT.2016.7474948" target="_blank" >http://dx.doi.org/10.1109/ICIT.2016.7474948</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/ICIT.2016.7474948" target="_blank" >10.1109/ICIT.2016.7474948</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Corrosion and mechanical properties of atomic layer deposited TiO2 Coatings on NiTi implants
Popis výsledku v původním jazyce
In the case of biomedical NiTi implants staying in the human body not only for a short time, a release of Ni ions from NiTi surface should be prevented because of the poisonous character of Ni. We modified the surface of NiTi plate samples by depositing TiO2 oxide using atomic layer deposition (ALD). We found excellent corrosion properties of the coated NiTi samples, implying that ALD has great potential in depositing protection or biocompatible layer for 3D printing. Additionally, using a mechanical tester installed in the sample chamber of a scanning electron microscope, we were not able to see any tensile-loadinduced microcracks in the thin atomic layer deposited TiO2 coating. Whereas the TiO2 layers grown thermally by oxidizing NiTi at temperature around 450 degrees C contained densely distributed microcracks during loading.
Název v anglickém jazyce
Corrosion and mechanical properties of atomic layer deposited TiO2 Coatings on NiTi implants
Popis výsledku anglicky
In the case of biomedical NiTi implants staying in the human body not only for a short time, a release of Ni ions from NiTi surface should be prevented because of the poisonous character of Ni. We modified the surface of NiTi plate samples by depositing TiO2 oxide using atomic layer deposition (ALD). We found excellent corrosion properties of the coated NiTi samples, implying that ALD has great potential in depositing protection or biocompatible layer for 3D printing. Additionally, using a mechanical tester installed in the sample chamber of a scanning electron microscope, we were not able to see any tensile-loadinduced microcracks in the thin atomic layer deposited TiO2 coating. Whereas the TiO2 layers grown thermally by oxidizing NiTi at temperature around 450 degrees C contained densely distributed microcracks during loading.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
10302 - Condensed matter physics (including formerly solid state physics, supercond.)
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of IEEE International Conference on Industrial Technology (ICIT) 2016
ISBN
978-1-4673-8075-1
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
1328-1332
Název nakladatele
IEEE
Místo vydání
New York
Místo konání akce
Taipei
Datum konání akce
14. 3. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000386327700206