Implementation of a crossed-slit system for fast alignment of sealed polycapillary X-ray optics
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F20%3A00539167" target="_blank" >RIV/68378271:_____/20:00539167 - isvavai.cz</a>
Výsledek na webu
<a href="http://hdl.handle.net/11104/0316907" target="_blank" >http://hdl.handle.net/11104/0316907</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1107/S1600577520012217" target="_blank" >10.1107/S1600577520012217</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Implementation of a crossed-slit system for fast alignment of sealed polycapillary X-ray optics
Popis výsledku v původním jazyce
A new modification of a table-top laser-driven water-jet plasma X-ray source has been successfully implemented and commissioned at the Extreme Light Infrastructure (ELI) Beamlines user facility. In order to preserve the broadband nature of the source for spectroscopic experiments, a polycapillary lens was initially chosen as the focusing element. Generally, polycapillary X-ray optics have a narrow photon acceptance angle and small field of view, making alignment complicated and time-consuming. This contribution demonstrates a straightforward, reliable and reproducible procedure for aligning polycapillary focusing optics with broadband X-rays.
Název v anglickém jazyce
Implementation of a crossed-slit system for fast alignment of sealed polycapillary X-ray optics
Popis výsledku anglicky
A new modification of a table-top laser-driven water-jet plasma X-ray source has been successfully implemented and commissioned at the Extreme Light Infrastructure (ELI) Beamlines user facility. In order to preserve the broadband nature of the source for spectroscopic experiments, a polycapillary lens was initially chosen as the focusing element. Generally, polycapillary X-ray optics have a narrow photon acceptance angle and small field of view, making alignment complicated and time-consuming. This contribution demonstrates a straightforward, reliable and reproducible procedure for aligning polycapillary focusing optics with broadband X-rays.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10301 - Atomic, molecular and chemical physics (physics of atoms and molecules including collision, interaction with radiation, magnetic resonances, Mössbauer effect)
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Synchrotron Radiation
ISSN
0909-0495
e-ISSN
—
Svazek periodika
27
Číslo periodika v rámci svazku
Nov
Stát vydavatele periodika
DK - Dánské království
Počet stran výsledku
4
Strana od-do
1730-1733
Kód UT WoS článku
000588645400032
EID výsledku v databázi Scopus
2-s2.0-85095677027