Popis výsledku v původním jazyce
The paper is focused on possibilities of modern X-ray detectors and micro-focus X-ray source for investigation of early degradation processes of acetabular implants. To simulate the most adverse activity (downstairs walking) a hip joint simulator was developed. The experimental setup was designed for cyclic loading of polyethylene acetabular cup implanted into the human pelvic bone and fixed by commercial polymethyl methacrylate bone cement. To predict the bone degradation numerical analysis of detailedthree-dimensional model of the acetabular cup and the cement mantle implanted in a bone block was performed. Using large area flat panel detector and microfocus X-ray source it is possible to investigate micro-damage propagation and detect early defectin the bone-implant interface.
Název v anglickém jazyce
Early Defect Detection of Acetabular Implants
Popis výsledku anglicky
The paper is focused on possibilities of modern X-ray detectors and micro-focus X-ray source for investigation of early degradation processes of acetabular implants. To simulate the most adverse activity (downstairs walking) a hip joint simulator was developed. The experimental setup was designed for cyclic loading of polyethylene acetabular cup implanted into the human pelvic bone and fixed by commercial polymethyl methacrylate bone cement. To predict the bone degradation numerical analysis of detailedthree-dimensional model of the acetabular cup and the cement mantle implanted in a bone block was performed. Using large area flat panel detector and microfocus X-ray source it is possible to investigate micro-damage propagation and detect early defectin the bone-implant interface.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FI - Traumatologie a ortopedie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GAP105%2F10%2F2305" target="_blank" >GAP105/10/2305: Morfometrické a mechanické vlastnosti trabekulární kosti zjištěné pomocí metod mikromechaniky a numerického modelování</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Engineering Mechanics 2012 - CD ROM
ISBN
978-80-86246-40-6
ISSN
—
e-ISSN
—
Počet stran výsledku
10
Strana od-do
825-834