Design and use of novel compression device for microtomography under applied load
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378297%3A_____%2F12%3A00384441" target="_blank" >RIV/68378297:_____/12:00384441 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.engmech.cz/2012/proceedings/" target="_blank" >http://www.engmech.cz/2012/proceedings/</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design and use of novel compression device for microtomography under applied load
Popis výsledku v původním jazyce
This paper deals with modification and usage of custom-designed compression device, that allows real time X-ray tomography scanning of specimen under applied pressure. In this case microtomography is used to obtain data required to determine specimens morphology and to develop 3D material model (especially for cellular materials such as bones, metal foams and quasi-brittle materials or particle composites such as concrete or cementitious composites). Important design changes were made in the existing device frame to increase its load capabilities, stiffness and to accomodate a larger specimen. Finally device displacement measurements were conducted and calibration experiment was carried out.
Název v anglickém jazyce
Design and use of novel compression device for microtomography under applied load
Popis výsledku anglicky
This paper deals with modification and usage of custom-designed compression device, that allows real time X-ray tomography scanning of specimen under applied pressure. In this case microtomography is used to obtain data required to determine specimens morphology and to develop 3D material model (especially for cellular materials such as bones, metal foams and quasi-brittle materials or particle composites such as concrete or cementitious composites). Important design changes were made in the existing device frame to increase its load capabilities, stiffness and to accomodate a larger specimen. Finally device displacement measurements were conducted and calibration experiment was carried out.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JI - Kompositní materiály
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GAP105%2F10%2F2305" target="_blank" >GAP105/10/2305: Morfometrické a mechanické vlastnosti trabekulární kosti zjištěné pomocí metod mikromechaniky a numerického modelování</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Engineering Mechanics 2012
ISBN
978-80-86246-40-6
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
239-244
Název nakladatele
ITAM AS CR, v. v. i.
Místo vydání
Prague
Místo konání akce
Svratka
Datum konání akce
14. 5. 2012
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—