Comparation of mechanical responses of dental implants manufactured by 3D printing and machining
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21110%2F17%3A00312969" target="_blank" >RIV/68407700:21110/17:00312969 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21220/17:00312969
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparation of mechanical responses of dental implants manufactured by 3D printing and machining
Popis výsledku v původním jazyce
This paper deals with the basic comparison of dental implant stems behavior during static loading tests. The implants used for the tests were manufactured by the mechanical machining method with use of the standard alloy Ti6Al4V and experimental alloy Ti35Nb6Ta. Another implant used for loading tests was produced by the method of 3D printing from the standard alloy Ti6Al4V. The shape solutions of used implants are based on patent files no. CZ 306456 and CZ 306457. The method of loading test is based on valid legislative, which is necessary for product launch on the market and it is managed by ČSN EN ISO 140801 (Dentistry - Implants - Dynamic loading test for endosseous dental implants).
Název v anglickém jazyce
Comparation of mechanical responses of dental implants manufactured by 3D printing and machining
Popis výsledku anglicky
This paper deals with the basic comparison of dental implant stems behavior during static loading tests. The implants used for the tests were manufactured by the mechanical machining method with use of the standard alloy Ti6Al4V and experimental alloy Ti35Nb6Ta. Another implant used for loading tests was produced by the method of 3D printing from the standard alloy Ti6Al4V. The shape solutions of used implants are based on patent files no. CZ 306456 and CZ 306457. The method of loading test is based on valid legislative, which is necessary for product launch on the market and it is managed by ČSN EN ISO 140801 (Dentistry - Implants - Dynamic loading test for endosseous dental implants).
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20501 - Materials engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Experimental Stress Analysis 2017
ISBN
978-80-553-3167-6
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
45-49
Název nakladatele
Technical University of Kosice
Místo vydání
Košice
Místo konání akce
Nový Smokovec
Datum konání akce
30. 5. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—