NUMERICAL COMPARISON OF TRANSGINGIVAL AND SUBGINGIVAL DENTAL IMPLANTS IN REGARD TO THEIR STRESS DISTRIBUTIONS
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21110%2F21%3A00352067" target="_blank" >RIV/68407700:21110/21:00352067 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.14311/APP.2021.30.0081" target="_blank" >http://dx.doi.org/10.14311/APP.2021.30.0081</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.14311/APP.2021.30.0081" target="_blank" >10.14311/APP.2021.30.0081</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
NUMERICAL COMPARISON OF TRANSGINGIVAL AND SUBGINGIVAL DENTAL IMPLANTS IN REGARD TO THEIR STRESS DISTRIBUTIONS
Popis výsledku v původním jazyce
Most modern dental implants differentiate in regard to the fixation of the abutment into two main categories - the external or internal hexagon or octagon. We performed mechanical tests according to the CSN EN ISO 14801 standard on a dental implant variant using the external hex. We found that failure of all implant specimens occured below the screw head. To improve the current geometry, we performed numerical analysis of an alternative variant (internal hex) and compared it with analysis of the current geometry (external hex). It was found that the stress distribution of the variant with internal hex is preferable to the old variant. Although extreme values of shear stress in the corresponding plane of loading are higher, they do not concentrate below the screw head, where the screw itself is thinner and more prone to breaking. Therefore, it seems that the new variant of the dental implant is stronger, which is still to be proven by mechanical tests.
Název v anglickém jazyce
NUMERICAL COMPARISON OF TRANSGINGIVAL AND SUBGINGIVAL DENTAL IMPLANTS IN REGARD TO THEIR STRESS DISTRIBUTIONS
Popis výsledku anglicky
Most modern dental implants differentiate in regard to the fixation of the abutment into two main categories - the external or internal hexagon or octagon. We performed mechanical tests according to the CSN EN ISO 14801 standard on a dental implant variant using the external hex. We found that failure of all implant specimens occured below the screw head. To improve the current geometry, we performed numerical analysis of an alternative variant (internal hex) and compared it with analysis of the current geometry (external hex). It was found that the stress distribution of the variant with internal hex is preferable to the old variant. Although extreme values of shear stress in the corresponding plane of loading are higher, they do not concentrate below the screw head, where the screw itself is thinner and more prone to breaking. Therefore, it seems that the new variant of the dental implant is stronger, which is still to be proven by mechanical tests.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
30404 - Biomaterials (as related to medical implants, devices, sensors)
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
NMM 2020 Nano & Macro Mechanics
ISBN
978-80-01-06840-3
ISSN
—
e-ISSN
2336-5382
Počet stran výsledku
6
Strana od-do
81-86
Název nakladatele
České vysoké učení technické v Praze
Místo vydání
Praha
Místo konání akce
Praha
Datum konání akce
17. 9. 2020
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—