Energy and Dose Rate Dependence of BANG-2 Polymer-Gel Dosimeter
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21220%2F01%3A04068256" target="_blank" >RIV/68407700:21220/01:04068256 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Energy and Dose Rate Dependence of BANG-2 Polymer-Gel Dosimeter
Popis výsledku v původním jazyce
The purpose of this study was to evaluate dependence of BANG-2 polymer-gel dosimeter sensitivity on different photon and electron energies as well as on different mean dose rates expressed as repetition rates for standard clinically used linear accelerator. The sensitivity of dosimeter was represented by the slope of calibration curve in the linear region measured for each modality. Calibration curve (in the linear region) based on five dosimeters (four irradiated and one background) was obtained for each photon and electron energy and different repetition rates. Dosimeter sensitivity dependence on energy was studied for 4, 6 and 18 MV X ray photons and for nominal electron energies 9, 12, 16, 20 MeV. Dosimeter sensitivity dependence on mean dose ratewas separately studied for electron and photon beams with the use of repetition rates 80, 160, 240, 320, 400 MU.min-1. Evaluation of dosimeters was performed on Siemens MAGNETOM EXPERT 1T scanner in the head coil. Multi-echo sequence wit.
Název v anglickém jazyce
Energy and Dose Rate Dependence of BANG-2 Polymer-Gel Dosimeter
Popis výsledku anglicky
The purpose of this study was to evaluate dependence of BANG-2 polymer-gel dosimeter sensitivity on different photon and electron energies as well as on different mean dose rates expressed as repetition rates for standard clinically used linear accelerator. The sensitivity of dosimeter was represented by the slope of calibration curve in the linear region measured for each modality. Calibration curve (in the linear region) based on five dosimeters (four irradiated and one background) was obtained for each photon and electron energy and different repetition rates. Dosimeter sensitivity dependence on energy was studied for 4, 6 and 18 MV X ray photons and for nominal electron energies 9, 12, 16, 20 MeV. Dosimeter sensitivity dependence on mean dose ratewas separately studied for electron and photon beams with the use of repetition rates 80, 160, 240, 320, 400 MU.min-1. Evaluation of dosimeters was performed on Siemens MAGNETOM EXPERT 1T scanner in the head coil. Multi-echo sequence wit.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2001
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Medical Physics
ISSN
0094-2405
e-ISSN
—
Svazek periodika
28
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
2379-2386
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—