Nové řešení uchycení šlachy pro stanovení její deformace bez použití extenzometru
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21220%2F08%3A02147678" target="_blank" >RIV/68407700:21220/08:02147678 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The New Solution of Clamping Problem in Tendon Strain Estimation Without Using of an Extensometer
Popis výsledku v původním jazyce
In the present study the in vitro tensile properties of the porcine extensor digitorum profundus tendons were obtained. For a measurement the testing system MTS Mini Bionix 858.2 was used. Data were obtained from displacement of jaws and with an extensometer. Data from displacement of jaws are influenced of slip specimens from jaws. The problem of the tendon sidle out of the jaws was solved. For the better fixation of tendon in jaws during tensile tests the new profile of jaws component were developed.Next innovation in tendon fixing comes with the cooling the tendon ends by liquid nitrogen before clamping. With the measuring setup with frozen ends and jagged rounded profile of jaws can be obtained satisfactory strain data without using of an extensometer.
Název v anglickém jazyce
The New Solution of Clamping Problem in Tendon Strain Estimation Without Using of an Extensometer
Popis výsledku anglicky
In the present study the in vitro tensile properties of the porcine extensor digitorum profundus tendons were obtained. For a measurement the testing system MTS Mini Bionix 858.2 was used. Data were obtained from displacement of jaws and with an extensometer. Data from displacement of jaws are influenced of slip specimens from jaws. The problem of the tendon sidle out of the jaws was solved. For the better fixation of tendon in jaws during tensile tests the new profile of jaws component were developed.Next innovation in tendon fixing comes with the cooling the tendon ends by liquid nitrogen before clamping. With the measuring setup with frozen ends and jagged rounded profile of jaws can be obtained satisfactory strain data without using of an extensometer.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JJ - Ostatní materiály
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP106%2F06%2FP304" target="_blank" >GP106/06/P304: Měření mechanických vlastností biologického materiálu optickými vláknovými senzory.</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Mechanica Slovaca
ISSN
1335-2393
e-ISSN
—
Svazek periodika
12
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
SK - Slovenská republika
Počet stran výsledku
8
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—