Popis výsledku v původním jazyce
The study deals with a stress analysis of the total wrist joint replacement. The replacement consists of eight parts grouped into radial and carpal components. Replacement parts are made of Ti6Al4V alloy except articulating parts, i.e. the spherical headmade of CoCrMo alloy and the joint cup insert made of ultra high molecular weight polyethylene. The aim of the analysis is to verify strength of the replacement so the case of fall on hands is simulated. The analysis is based on the finite element method and is performed by means of ABAQUS software package. Elastic-plastic material models are applied. Computations of the stresses within radial and carpal components are proceeded separately. Components are loaded by force located in the joint centre, coming from published experimental studies. Results of computations prove strength of the replacement, stress within components and contact pressure on articulating surfaces do not exceed limit values.
Název v anglickém jazyce
Computational Stress Analysis of the Total Wrist Replacement Medin
Popis výsledku anglicky
The study deals with a stress analysis of the total wrist joint replacement. The replacement consists of eight parts grouped into radial and carpal components. Replacement parts are made of Ti6Al4V alloy except articulating parts, i.e. the spherical headmade of CoCrMo alloy and the joint cup insert made of ultra high molecular weight polyethylene. The aim of the analysis is to verify strength of the replacement so the case of fall on hands is simulated. The analysis is based on the finite element method and is performed by means of ABAQUS software package. Elastic-plastic material models are applied. Computations of the stresses within radial and carpal components are proceeded separately. Components are loaded by force located in the joint centre, coming from published experimental studies. Results of computations prove strength of the replacement, stress within components and contact pressure on articulating surfaces do not exceed limit values.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BO - Biofyzika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/FI-IM5%2F139" target="_blank" >FI-IM5/139: *Řešení biomechanických poměrů kloubních spojení ruky a nohy - vývoj náhrady kloubu zápěstí a náhrad malých kloubů ruky a nohy</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Human Biomechanics 2010
ISBN
978-80-7372-648-5
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
—
Název nakladatele
Technical University of Liberec
Místo vydání
Liberec
Místo konání akce
Sychrov
Datum konání akce
4. 10. 2010
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—