A Weak Magnetic Field Measurement System Using Micro-Fluxgate Sensors and Delta-Sigma Interface
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F03%3A03087195" target="_blank" >RIV/68407700:21230/03:03087195 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A Weak Magnetic Field Measurement System Using Micro-Fluxgate Sensors and Delta-Sigma Interface
Popis výsledku v původním jazyce
This paper presents a weak magnetic field measurement system using micro-fluxgate (FG) sensors and a sensor signal processing technique using the delta-sigma modulation in the negative feedback loop. The feedback of the lowpass filtered bitstream outputof a delta-sigma modulator to the magnetic field improves system linearity, hysteresis, and stability. In spite of the fact that the second-order delta-sigma modulator is used, the third-order noise shaping can be obtained in the digital output bit-stream by the use of an integrator in the loop. This improves the SNR of the digital output. The measured noise of the implemented system meets the measured noise of the FG sensing element itself. The weak magnetic field in the range of the Earth's magnetic field is successfully measured. The nonlinearity error is less then 0.4 % in the range of +/- 100 uT.
Název v anglickém jazyce
A Weak Magnetic Field Measurement System Using Micro-Fluxgate Sensors and Delta-Sigma Interface
Popis výsledku anglicky
This paper presents a weak magnetic field measurement system using micro-fluxgate (FG) sensors and a sensor signal processing technique using the delta-sigma modulation in the negative feedback loop. The feedback of the lowpass filtered bitstream outputof a delta-sigma modulator to the magnetic field improves system linearity, hysteresis, and stability. In spite of the fact that the second-order delta-sigma modulator is used, the third-order noise shaping can be obtained in the digital output bit-stream by the use of an integrator in the loop. This improves the SNR of the digital output. The measured noise of the implemented system meets the measured noise of the FG sensing element itself. The weak magnetic field in the range of the Earth's magnetic field is successfully measured. The nonlinearity error is less then 0.4 % in the range of +/- 100 uT.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2003
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Instrumentation and Measurement
ISSN
0018-9456
e-ISSN
—
Svazek periodika
52
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
103-110
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—