Design and Realization of Hardware for Measurement of Hemodynamic Parameters
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F13%3A00193587" target="_blank" >RIV/68407700:21230/13:00193587 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design and Realization of Hardware for Measurement of Hemodynamic Parameters
Popis výsledku v původním jazyce
This paper presents the design and realization of the hardware for measurement of hemodynamic parameters. The aim of the work has been to develop the hardware for sensing of biological signals such as the oscillometric pulsations in arm cuff and in wristcuff, the photoplethysmography (PPG) signal and the electrocardiography (ECG) signal. The selection of signals allows a determination of selected hemodynamic parameters, which are important for primary screening of atherosclerosis, for example the pulsewave velocity (PWV) or the arterial stiffness index (ASI). The hardware is designed for measuring biological signals both during the increasing of pressure in the arm cuff and during the decreasing of the pressure. The measuring of oscillometric pulsations in the cuffs during the increasing of the pressure eliminates the artefacts caused by deformation of the arm tissues. The design and realization of the hardware is described in this paper. The paper includes the example of the measure
Název v anglickém jazyce
Design and Realization of Hardware for Measurement of Hemodynamic Parameters
Popis výsledku anglicky
This paper presents the design and realization of the hardware for measurement of hemodynamic parameters. The aim of the work has been to develop the hardware for sensing of biological signals such as the oscillometric pulsations in arm cuff and in wristcuff, the photoplethysmography (PPG) signal and the electrocardiography (ECG) signal. The selection of signals allows a determination of selected hemodynamic parameters, which are important for primary screening of atherosclerosis, for example the pulsewave velocity (PWV) or the arterial stiffness index (ASI). The hardware is designed for measuring biological signals both during the increasing of pressure in the arm cuff and during the decreasing of the pressure. The measuring of oscillometric pulsations in the cuffs during the increasing of the pressure eliminates the artefacts caused by deformation of the arm tissues. The design and realization of the hardware is described in this paper. The paper includes the example of the measure
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
IFMBE Proceedings: World Congress on Medical Physics and Biomedical Engineering
ISBN
978-3-642-29304-7
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1420-1423