E-field distribution measurement system for hyperthermia applicators with LED sensor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F15%3A00233721" target="_blank" >RIV/68407700:21230/15:00233721 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/COMITE.2015.7120227" target="_blank" >http://dx.doi.org/10.1109/COMITE.2015.7120227</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/COMITE.2015.7120227" target="_blank" >10.1109/COMITE.2015.7120227</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
E-field distribution measurement system for hyperthermia applicators with LED sensor
Popis výsledku v původním jazyce
In this contribution we present a first version of a simple and fast system for E-field and SAR distribution measurement of hyperthermia applicators. The system consists of five main parts. These are 3D positioning system, light emitting diode (LED) sensor of electrical field (E-field), aquarium with saline solution, digital camera and image processing algorithm. LED E-field sensor is moved in front of the aperture of hyperthermia applicator while the digital camera takes pictures of the scene in appropriate cross section of the applicator. Digital image processing is then performed in order to extract intensity of light emitted from the sensor and reconstruct the final distribution of E-field. Results from this measuring procedure are compared with the results from numerical simulation.
Název v anglickém jazyce
E-field distribution measurement system for hyperthermia applicators with LED sensor
Popis výsledku anglicky
In this contribution we present a first version of a simple and fast system for E-field and SAR distribution measurement of hyperthermia applicators. The system consists of five main parts. These are 3D positioning system, light emitting diode (LED) sensor of electrical field (E-field), aquarium with saline solution, digital camera and image processing algorithm. LED E-field sensor is moved in front of the aperture of hyperthermia applicator while the digital camera takes pictures of the scene in appropriate cross section of the applicator. Digital image processing is then performed in order to extract intensity of light emitted from the sensor and reconstruct the final distribution of E-field. Results from this measuring procedure are compared with the results from numerical simulation.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of 14th Conference on Microwave Techniques, COMITE 2015
ISBN
9781479981212
ISSN
—
e-ISSN
—
Počet stran výsledku
3
Strana od-do
30-32
Název nakladatele
Institute of Electrical and Electronics Engineers Inc.
Místo vydání
Budapešť
Místo konání akce
Pardubice
Datum konání akce
21. 4. 2015
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—