Off-Center Error Correction of AMR Yokeless Current Transducer
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F17%3A00314512" target="_blank" >RIV/68407700:21230/17:00314512 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.hindawi.com/journals/js/2017/6057634/" target="_blank" >https://www.hindawi.com/journals/js/2017/6057634/</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1155/2017/6057634" target="_blank" >10.1155/2017/6057634</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Off-Center Error Correction of AMR Yokeless Current Transducer
Popis výsledku v původním jazyce
We present a method of calibration and error correction of the AMR yokeless current transducer consisting of a circular array of eight anisotropic magnetoresistors (AMR) with one feedback compensation loop. The main sources of errors are the nonidentical parameters of AMR sensors and off-center position of the measured current. It is well known that AMR sensors from the same batch have 2% spread of the sensitivity; we found that the variation of the factor of the internal compensation coil is the same. We developed a novel calibration process using the readings of individual residual uncompensated voltages of the AMRs. The position of the current inside the measurement hole is estimated from the individual voltages considering the influence of external DC magnetic field such as the Earth’s field. During the calibration phase, the sensor outputs are measured for several positions of the current conductor inside the measuring hole. As a result of calibration the lookup table of error corrections is calculated and stored in the memory, and then these values are used for the correction during the measurement of the unknown current. This procedure reduces the off-center error from 0.4% to 0.06%.
Název v anglickém jazyce
Off-Center Error Correction of AMR Yokeless Current Transducer
Popis výsledku anglicky
We present a method of calibration and error correction of the AMR yokeless current transducer consisting of a circular array of eight anisotropic magnetoresistors (AMR) with one feedback compensation loop. The main sources of errors are the nonidentical parameters of AMR sensors and off-center position of the measured current. It is well known that AMR sensors from the same batch have 2% spread of the sensitivity; we found that the variation of the factor of the internal compensation coil is the same. We developed a novel calibration process using the readings of individual residual uncompensated voltages of the AMRs. The position of the current inside the measurement hole is estimated from the individual voltages considering the influence of external DC magnetic field such as the Earth’s field. During the calibration phase, the sensor outputs are measured for several positions of the current conductor inside the measuring hole. As a result of calibration the lookup table of error corrections is calculated and stored in the memory, and then these values are used for the correction during the measurement of the unknown current. This procedure reduces the off-center error from 0.4% to 0.06%.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/GA17-19877S" target="_blank" >GA17-19877S: Nové metody měření elektrických proudů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Sensors
ISSN
1687-725X
e-ISSN
1687-7268
Svazek periodika
2017
Číslo periodika v rámci svazku
October
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
—
Kód UT WoS článku
000414099100001
EID výsledku v databázi Scopus
2-s2.0-85042272088