Popis výsledku v původním jazyce
In this paper, we propose a novel idea of a diagnostic applicator composed of an array of 8 waveguides filled with distilled water, which can be used for non-invasive temperature measurement based on microwave tomography during cancer treatment using microwave hyperthermia. The waveguide design procedure and simulation results from SEMCAD X are presented. To verify the impedance matching of the applicator, scattering matrix measurements were performed with a muscle tissue agar phantom. Measurements with dielectric inhomogeneity in the agar phantom region are presented to demonstrate the feasibility of using our design for further developments in this area.
Název v anglickém jazyce
Waveguide Array Applicator for Microwave Medical Imaging
Popis výsledku anglicky
In this paper, we propose a novel idea of a diagnostic applicator composed of an array of 8 waveguides filled with distilled water, which can be used for non-invasive temperature measurement based on microwave tomography during cancer treatment using microwave hyperthermia. The waveguide design procedure and simulation results from SEMCAD X are presented. To verify the impedance matching of the applicator, scattering matrix measurements were performed with a muscle tissue agar phantom. Measurements with dielectric inhomogeneity in the agar phantom region are presented to demonstrate the feasibility of using our design for further developments in this area.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 52nd European Microwave Conference
ISBN
9782874870699
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
187-190
Název nakladatele
European Microwave Association (EuMA)
Místo vydání
Louvain-la-Neuve
Místo konání akce
Milano
Datum konání akce
27. 9. 2022