Timepix3: Temperature Influence on Radiation Energy Measurement with Si Sensor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F23%3A00363682" target="_blank" >RIV/68407700:21230/23:00363682 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.3390/s23042201" target="_blank" >https://doi.org/10.3390/s23042201</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/s23042201" target="_blank" >10.3390/s23042201</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Timepix3: Temperature Influence on Radiation Energy Measurement with Si Sensor
Popis výsledku v původním jazyce
The Timepix3 readout ASIC chip is a hybrid pixelated radiation detector, designed at CERN, which contains a 256 pxx256 px matrix. Each of the 65,536 radiation-sensitive pixels can record an incoming particle, its energy deposition or time of arrival and measure them simultaneously. Since the detector is suitable for a wide range of applications from particle physics, national security and medicine to space science, it can be used in a wide range of temperatures. Until now, it has to be calibrated every time to the operating point of the application. This paper studies the possibility of energy measurement with Timepix3 equipped with a 500 µm thick silicon sensor and MiniPIX readout interface in the temperatures between 10 °C and 70 °C with only one calibration. The detector has been irradiated by X-ray fluorescence photons in the energy range from 8 keV to 57 keV, and 31 keV to 81 keV photons from the 133Ba radioactive source. A deviation of 5% in apparent energy value may occur for a 10 °C change in temperature from the reference point, but, with the next temperature change, it can reach up to -30%. Moreover, Barium photons with an energy of 81 keV appear as deposited energy of only 55 keV at a detector temperature of 70 °C. An original compensation method that reduces the relative measurement error from -30% to less than 1% is presented in this paper.
Název v anglickém jazyce
Timepix3: Temperature Influence on Radiation Energy Measurement with Si Sensor
Popis výsledku anglicky
The Timepix3 readout ASIC chip is a hybrid pixelated radiation detector, designed at CERN, which contains a 256 pxx256 px matrix. Each of the 65,536 radiation-sensitive pixels can record an incoming particle, its energy deposition or time of arrival and measure them simultaneously. Since the detector is suitable for a wide range of applications from particle physics, national security and medicine to space science, it can be used in a wide range of temperatures. Until now, it has to be calibrated every time to the operating point of the application. This paper studies the possibility of energy measurement with Timepix3 equipped with a 500 µm thick silicon sensor and MiniPIX readout interface in the temperatures between 10 °C and 70 °C with only one calibration. The detector has been irradiated by X-ray fluorescence photons in the energy range from 8 keV to 57 keV, and 31 keV to 81 keV photons from the 133Ba radioactive source. A deviation of 5% in apparent energy value may occur for a 10 °C change in temperature from the reference point, but, with the next temperature change, it can reach up to -30%. Moreover, Barium photons with an energy of 81 keV appear as deposited energy of only 55 keV at a detector temperature of 70 °C. An original compensation method that reduces the relative measurement error from -30% to less than 1% is presented in this paper.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_019%2F0000778" target="_blank" >EF16_019/0000778: Centrum pokročilých aplikovaných přírodních věd</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2023
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Sensors
ISSN
1424-8220
e-ISSN
1424-8220
Svazek periodika
23
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
14
Strana od-do
—
Kód UT WoS článku
000942406500001
EID výsledku v databázi Scopus
2-s2.0-85149168457