Nonlinear Distortion of Microphone in Probe for Otoacoustic Emission Measurements
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F24%3A00382990" target="_blank" >RIV/68407700:21230/24:00382990 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.czakustika.cz/misc/aklisty/27-14.pdf" target="_blank" >https://www.czakustika.cz/misc/aklisty/27-14.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Nonlinear Distortion of Microphone in Probe for Otoacoustic Emission Measurements
Popis výsledku v původním jazyce
Nonlinearities in otoacoustic emission (OAE) measurement probes can significantly impact the accuracy of distortion-product otoacoustic emissions (DPOAE) results. This paper investigates the nonlinear behavior of condenser microphones used in these probes. Due to the lack of a perfectly linear reference source, we employ harmonic correction methods found in the literature to address the nonlinearities. We measure the second and third harmonics produced by the microphone, estimate key parameters of the nonlinear microphone model from these measurements, and use this model to predict the nonlinear behavior of the microphone in two different measurement scenarios. In the first scenario, without a low-frequency bias tone, the nonlinear distortion produced by the microphone is negligible. In the second scenario, with a low-frequency bias tone, the nonlinear distortion becomes significant. The findings demonstrate that we are able to measure the microphone’s nonlinearities and estimate key model parameters to predict the significance of these nonlinearities in different measurement scenarios.
Název v anglickém jazyce
Nonlinear Distortion of Microphone in Probe for Otoacoustic Emission Measurements
Popis výsledku anglicky
Nonlinearities in otoacoustic emission (OAE) measurement probes can significantly impact the accuracy of distortion-product otoacoustic emissions (DPOAE) results. This paper investigates the nonlinear behavior of condenser microphones used in these probes. Due to the lack of a perfectly linear reference source, we employ harmonic correction methods found in the literature to address the nonlinearities. We measure the second and third harmonics produced by the microphone, estimate key parameters of the nonlinear microphone model from these measurements, and use this model to predict the nonlinear behavior of the microphone in two different measurement scenarios. In the first scenario, without a low-frequency bias tone, the nonlinear distortion produced by the microphone is negligible. In the second scenario, with a low-frequency bias tone, the nonlinear distortion becomes significant. The findings demonstrate that we are able to measure the microphone’s nonlinearities and estimate key model parameters to predict the significance of these nonlinearities in different measurement scenarios.
Klasifikace
Druh
J<sub>ost</sub> - Ostatní články v recenzovaných periodicích
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/GC23-07621J" target="_blank" >GC23-07621J: Otoakustické emise v normální kochlee a při přetlaku v endolymfatickém prostoru kochley: modelování a experimenty</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Akustické listy
ISSN
1212-4702
e-ISSN
—
Svazek periodika
27
Číslo periodika v rámci svazku
1-4
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
4
Strana od-do
3-6