Study of Temperature Influence on Timepix2 Energy Measurements
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F25%3A00388247" target="_blank" >RIV/68407700:21230/25:00388247 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21670/25:00388247
Výsledek na webu
<a href="https://doi.org/10.1088/1748-0221/20/11/C11003" target="_blank" >https://doi.org/10.1088/1748-0221/20/11/C11003</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/20/11/C11003" target="_blank" >10.1088/1748-0221/20/11/C11003</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Study of Temperature Influence on Timepix2 Energy Measurements
Popis výsledku v původním jazyce
In this study, we examine the temperature dependence of a Timepix2 (v1) detector (256x 256 pixels, 55 μm pixel pitch) coupled to a 500 μm thick silicon sensor. Measurements were performed with γ-photons 59.5 keV and α-particles 5.5 MeV in vacuum over a stabilized temperature range of -20 °C to 80 °C. Two different readout systems were used to verify that the response of the Timepix2 detector was not affected by the readout electronics. In both normal and adaptive gain modes, the detector demonstrated excellent stability in Time-over-Threshold (ToT) based energy measurements, with relative deviations of peak positions remaining below 4% across the temperature range from -20 °C to 80 °C. Digital-to-Analog Converter (DAC) voltage measurements revealed a significant temperature dependence of the DAC Ikrum, which compensates for threshold shifts and enables accurate ToT readings. In adaptive gain mode, a systematic shift of approximately 0.6% per 10 °C was observed in absolute ToT accuracy.
Název v anglickém jazyce
Study of Temperature Influence on Timepix2 Energy Measurements
Popis výsledku anglicky
In this study, we examine the temperature dependence of a Timepix2 (v1) detector (256x 256 pixels, 55 μm pixel pitch) coupled to a 500 μm thick silicon sensor. Measurements were performed with γ-photons 59.5 keV and α-particles 5.5 MeV in vacuum over a stabilized temperature range of -20 °C to 80 °C. Two different readout systems were used to verify that the response of the Timepix2 detector was not affected by the readout electronics. In both normal and adaptive gain modes, the detector demonstrated excellent stability in Time-over-Threshold (ToT) based energy measurements, with relative deviations of peak positions remaining below 4% across the temperature range from -20 °C to 80 °C. Digital-to-Analog Converter (DAC) voltage measurements revealed a significant temperature dependence of the DAC Ikrum, which compensates for threshold shifts and enables accurate ToT readings. In adaptive gain mode, a systematic shift of approximately 0.6% per 10 °C was observed in absolute ToT accuracy.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/GM23-04869M" target="_blank" >GM23-04869M: Identifikace částic v experimentech fysiky vysokych energií a ve vesmíru s pokročilými detekčními systémy</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
1748-0221
Svazek periodika
20
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
10
Strana od-do
1-10
Kód UT WoS článku
001615483400001
EID výsledku v databázi Scopus
2-s2.0-105033813431