Contact Stress Analysis of Cemented Acetabular Implant
Popis výsledku v původním jazyce
The paper deals with contact stress analysis of pelvic bone after implantation of cemented acetabular component with imperfections of cemented layer. 3D high resolution FE model of pelvic bone was created from the sequence of Computer Tomography (CT) slices. The magnitude and character of the contact stress distribution in subchondral bone depends on the type of imperfections in cement layer. Several openings are often drilled into the bone to improve the fixation of the cemented acetabular cup into thepelvic bone. Boolean operations were used for inserting the spherical cemented acetabular implant. Two contacts were defined in the FE model. The first contact is defined between the cement layer and the pelvic bone and the second one is between the cement layer and the polyethylene cup. The model was loaded by quasistatic joint forces, representing the maximum value of stance phase of gait.
Název v anglickém jazyce
Contact Stress Analysis of Cemented Acetabular Implant
Popis výsledku anglicky
The paper deals with contact stress analysis of pelvic bone after implantation of cemented acetabular component with imperfections of cemented layer. 3D high resolution FE model of pelvic bone was created from the sequence of Computer Tomography (CT) slices. The magnitude and character of the contact stress distribution in subchondral bone depends on the type of imperfections in cement layer. Several openings are often drilled into the bone to improve the fixation of the cemented acetabular cup into thepelvic bone. Boolean operations were used for inserting the spherical cemented acetabular implant. Two contacts were defined in the FE model. The first contact is defined between the cement layer and the pelvic bone and the second one is between the cement layer and the polyethylene cup. The model was loaded by quasistatic joint forces, representing the maximum value of stance phase of gait.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
EI - Biotechnologie a bionika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA103%2F00%2F0831" target="_blank" >GA103/00/0831: Biomechanická analýza lokálních účinků soustavy endoprotéza - pánev</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Fourth International IASTED Conference on BIOMECHANICS
ISBN
0-88986-604-X
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
11-14
Název nakladatele
IASTED
Místo vydání
Gran Meliá Salinas, Lanzarote
Místo konání akce
Palma de Mallorca
Datum konání akce
27. 8. 2006
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—