Verifikace dávkových objemových histogramů ve stereotaktické radiochirurgii a radioterapii prostřednictvím polymerních gelových dozimetrů a magnetické resonance
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F06%3A04123280" target="_blank" >RIV/68407700:21340/06:04123280 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Verification of Dose Volume Histograms in Stereotactic Radiosurgery and Radiotherapy Using Polymer gel and MRI
Popis výsledku v původním jazyce
Polymer gel (PG) dosimetry seems to be a promising method for verification of complex radiotherapeutic treatment techniques particularly via dose volume histograms (DVH). Golden standard for PG dosimeter response read out is magnetic resonance imaging (MRI) by which the transversal relaxation rate (denoted R2) is indirectly measured. The relaxation rate is altered by the polymer structure formed in the irradiated gel. The increase in R2 is proportional to absorbed dose. In the experiment we focused on the methodology on dose volume histograms measurement via PG dosimetry and MRI for stereotactic radiosurgery and radiotherapy. The proposed methodology was preliminary tested on the Leksell gamma knife system (ELEKTA Instrument AB, Sweden) and consecutively we used it for verification of three treatment plans created in BrainLab treatment planning system (BrainLAB, Germany) in conjunction with 6MV linear accelarator Clinac 2100 C/D (Varian Medical Systems, USA).
Název v anglickém jazyce
Verification of Dose Volume Histograms in Stereotactic Radiosurgery and Radiotherapy Using Polymer gel and MRI
Popis výsledku anglicky
Polymer gel (PG) dosimetry seems to be a promising method for verification of complex radiotherapeutic treatment techniques particularly via dose volume histograms (DVH). Golden standard for PG dosimeter response read out is magnetic resonance imaging (MRI) by which the transversal relaxation rate (denoted R2) is indirectly measured. The relaxation rate is altered by the polymer structure formed in the irradiated gel. The increase in R2 is proportional to absorbed dose. In the experiment we focused on the methodology on dose volume histograms measurement via PG dosimetry and MRI for stereotactic radiosurgery and radiotherapy. The proposed methodology was preliminary tested on the Leksell gamma knife system (ELEKTA Instrument AB, Sweden) and consecutively we used it for verification of three treatment plans created in BrainLab treatment planning system (BrainLAB, Germany) in conjunction with 6MV linear accelarator Clinac 2100 C/D (Varian Medical Systems, USA).
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FD - Onkologie a hematologie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/ME%20835" target="_blank" >ME 835: Třídimenzionální dozimetrie s využitím experimentálních a výpočetních metod pro zjištění kvality moderních metod v radiační onkologii</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Sborník rozšířených abstraktů XXVIII DRO
ISBN
80-01-03575-1
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
300-303
Název nakladatele
Ediční středisko ČVUT
Místo vydání
Praha
Místo konání akce
Luhačovice
Datum konání akce
20. 11. 2006
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—