Performance evaluation of fast, high precision laser rangefinder electronics with a pulsed laser
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F11%3A00185996" target="_blank" >RIV/68407700:21340/11:00185996 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1117/12.912323" target="_blank" >http://dx.doi.org/10.1117/12.912323</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1117/12.912323" target="_blank" >10.1117/12.912323</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Performance evaluation of fast, high precision laser rangefinder electronics with a pulsed laser
Popis výsledku v původním jazyce
A laser rangefinder system based on a pulsed time of flight distance measurement technique was constructed and tested. The system can achieve a timing resolution better than 15 ps rms with up to 90 kHz repetition rate. The measurement range is from 0 psup to 1 second. The performance of the system was evaluated at short distances with a pulsed laboratory constructed laser. It is a mode-locked Nd:YAG laser generating 20 ps output pulses at the wavelength of 1064 nm with a repetition rate up to 100 Hz. The system is capable of measuring distances down to 0 m. The overall measurement accuracy is estimated to be 5 mm and will be verified by the experiment.
Název v anglickém jazyce
Performance evaluation of fast, high precision laser rangefinder electronics with a pulsed laser
Popis výsledku anglicky
A laser rangefinder system based on a pulsed time of flight distance measurement technique was constructed and tested. The system can achieve a timing resolution better than 15 ps rms with up to 90 kHz repetition rate. The measurement range is from 0 psup to 1 second. The performance of the system was evaluated at short distances with a pulsed laboratory constructed laser. It is a mode-locked Nd:YAG laser generating 20 ps output pulses at the wavelength of 1064 nm with a repetition rate up to 100 Hz. The system is capable of measuring distances down to 0 m. The overall measurement accuracy is estimated to be 5 mm and will be verified by the experiment.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F09%2F1741" target="_blank" >GA102/09/1741: Diodově buzený výkonový pikosekundový laserový systém</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of SPIE Vol. 8306 - Photonics, Devices, and Systems V
ISBN
978-0-8194-8953-1
ISSN
0277-786X
e-ISSN
—
Počet stran výsledku
7
Strana od-do
"83060D-1"-"83060D-7"
Název nakladatele
SPIE
Místo vydání
Bellingham
Místo konání akce
Praha
Datum konání akce
24. 8. 2011
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—