Verification of Co-60 Blood Irradiator Mechanical Engineering Design
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F14%3A00224379" target="_blank" >RIV/68407700:21340/14:00224379 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Verification of Co-60 Blood Irradiator Mechanical Engineering Design
Popis výsledku v původním jazyce
Aim of this work is to verify the mechanical engineering design of the blood irradiator (from view of radiation protection) and estimation of the field homogeneity in the irradiation chamber. To estimate the radiation field around the blood irradiator amodeling approach have been chosen and Monte Carlo code MCNPX has been employed. Fluence-to-dose conversion factor from ICRP-74 have been used for the annual dose equivalent evaluation. With this results, the shielding analysis has been performed in thecontext of occupation factor at the service staff work place and the current legislation requirements. (In-)homogeneity of the radiation field has been estimated on the basis of isodoses in the water phantom (imitate the irradiator object, e.g. blood unit) placed in the irradiation chamber. The results have been further discussed with regards to the additional optimization (cost, availability of a sufficient number of resources, etc.).
Název v anglickém jazyce
Verification of Co-60 Blood Irradiator Mechanical Engineering Design
Popis výsledku anglicky
Aim of this work is to verify the mechanical engineering design of the blood irradiator (from view of radiation protection) and estimation of the field homogeneity in the irradiation chamber. To estimate the radiation field around the blood irradiator amodeling approach have been chosen and Monte Carlo code MCNPX has been employed. Fluence-to-dose conversion factor from ICRP-74 have been used for the annual dose equivalent evaluation. With this results, the shielding analysis has been performed in thecontext of occupation factor at the service staff work place and the current legislation requirements. (In-)homogeneity of the radiation field has been estimated on the basis of isodoses in the water phantom (imitate the irradiator object, e.g. blood unit) placed in the irradiation chamber. The results have been further discussed with regards to the additional optimization (cost, availability of a sufficient number of resources, etc.).
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA02010372" target="_blank" >TA02010372: Technologie recyklace vysoceaktivních uzavřených zdrojů ionizujícího záření pro jejich následné použití v průmyslových a zdravotnických aplikacích.</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů